FDA Adverse Event Injury Summary report: N

PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM

MDR report key: 1905020 · Received November 22, 2010

Report

Report Number
2953144-2010-02916
Event Type
Injury
Date Received
November 22, 2010
Date of Event
October 6, 2010
Report Date
October 26, 2010
Manufacturer
AV-REDWOOD CITY
Product Code
MGB
PMA / PMN Number
P960043
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS RECEIVED. INVESTIGATION IS NOT YET COMPLETE.(B)(4):DURING PROCESSING OF THIS COMPLAINT, ATTEMPTS WERE MADE TO OBTAIN COMPLETE EVENT, PATIENT AND DEVICE INFORMATION.

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: EVALUATION OF THE RETURNED DEVICE FOUND THAT THE ANTERIOR CUFF AND LINK WERE ENGAGED TO THE ANTERIOR NEEDLE TIP. THE POSTERIOR CUFF, NEEDLE TIP AND MONOFILAMENT WERE NOT RETURNED WITH THE DEVICE. BASED ON THE INVESTIGATION FINDINGS, THE LINK WAS PULLED FROM THE POSTERIOR CUFF. THE LINK CONNECTS THE ANTERIOR NEEDLE TO THE SUTURE AND IF THE LINK DETACHES FROM THE CUFF, THE SUTURE WILL NOT BE PRESENT DURING PLUNGER WITHDRAWAL AND MAY APPEARED LIKE A CUFF MISSED. THE LINK DETACHMENT IS LIKELY THE RESULT OF RESISTANCE OR DRAG ENCOUNTERED DURING THE PROCEDURE. NO MANUFACTURING OR QUALITY ISSUE WAS DETECTED. A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS LOT DID NOT PRODUCE ANY FINDINGS RELEVANT TO THIS REPORT.

Description of Event or Problem · 1

IT WAS REPORTED THAT A PHYSICIAN TRAINED IN THE USE OF THE PROGLIDE DEVICE ATTEMPTED ARTERIOTOMY CLOSURE OF THE COMMON FEMORAL ARTERY AFTER A DIAGNOSTIC PROCEDURE. REPORTEDLY, WHEN THE NEEDLE PLUNGER WAS REMOVED, A CUFF MISS OCCURRED. THE PROGLIDE WAS REMOVED AND A SECOND PROGLIDE DEVICE WAS USED WITH THE SAME RESULTS. HEMOSTASIS WAS ACHIEVED USING A NON-ABBOTT DEVICE. THERE WERE NO REPORTED ADVERSE PATIENT EFFECTS. THOUGH REQUESTED, NO ADDITIONAL INFORMATION WAS PROVIDED.

Description of Event or Problem · 1

WHILE TRYING TO SET UP THE ENDOPLEGE SINUS CATHETER KIT AT THE BEGINNING OF THE CASE, WE WERE UNABLE TO ENGAGE THE SHEATH TO THE CATHETER TO TIGHTEN OR CONNECT IT. THE THREADED END ACTED LIKE IT WAS GOING ON BUT WOULD NOT TIGHTEN AND THE CATHETER WAS FREE TO MOVE BACK AND FORTH. WE HAD TO OPEN UP ANOTHER KIT TO USE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM SUTURE MEDIATED CLOSURE MGB AV-REDWOOD CITY 930076H

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention