FDA Adverse Event Malfunction Summary report: N

COLLEAGUE 3 CXE VOLUMETRICINFUSION PUMP

MDR report key: 1904994 · Received November 22, 2010

Report

Report Number
6000001-2010-05139
Event Type
Malfunction
Date Received
November 22, 2010
Date of Event
September 18, 2010
Report Date
September 23, 2010
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FRN
PMA / PMN Number
K041191
Removal / Correction Number
6000001-12/1/08-002-C
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). A SERVICE HISTORY REVIEW REVEALED THAT THIS DEVICE WAS PREVIOUSLY SERVICED FOR THE REPORTED CONDITION. BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED CONDITION. THE ROOT CAUSE INVESTIGATION IS IN PROGRESS THROUGH (B)(4).

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE HAS BEEN EVALUATED ONSITE. THE REPORTED CONDITION WAS CONFIRMED. THE ASSIGNABLE CAUSE HAS BEEN DETERMINED TO BE DEPLETED MAIN BATTERIES. THE MAIN BATTERIES AND HARNESS WERE REPLACED. A FOLLOW-UP MEDWATCH REPORT WILL BE SUBMITTED IF ADDITIONAL INFORMATION BECOMES AVAILABLE.

Description of Event or Problem · 1

ON (B)(6) 2010 THE BAXTER FIELD SERVICE TECHNICIAN SERVICED A COLLEAGUE INFUSION PUMP FOR A BATTERY ISSUE. DURING BAXTER'S REVIEW OF THE EVENT HISTORY ON (B)(6) 2010 FOR ANOTHER REPORT, IT WAS DISCOVERED THAT A BATTERY DEPLETED SET ALARM OCCURRED DURING INFUSION WHICH CAUSED AN INTERRUPTION DURING DELIVERY. THERE WAS NO ADVERSE EVENT, PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS REPORT. THERE IS NO FURTHER COMPLAINT INFORMATION AVAILABLE. THE USER INTERFACE MODULE MASTER SOFTWARE VERSION IS 6.13.92, CATEGORIZED AS REMEDIATED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COLLEAGUE 3 CXE VOLUMETRICINFUSION PUMP PUMP, INFUSION FRN BAXTER HEALTHCARE - SINGAPORE

Patients

Seq Age Sex Outcome Treatment
1