COLLEAGUE SINGLE CHANNEL VOLUMETRIC PUMP CE FRENCH
Report
- Report Number
- 6000001-2010-05046
- Event Type
- Malfunction
- Date Received
- November 22, 2010
- Date of Event
- August 25, 2010
- Report Date
- September 2, 2010
- Manufacturer
- BAXTER HEALTHCARE - SINGAPORE
- Product Code
- FRN
- Removal / Correction Number
- 6000001-12/1/08-002C
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- OTHER
Narratives
(B)(4). THIS DEVICE IS DISTRIBUTED OUTSIDE OF THE UNITED STATES (US); THEREFORE, IT DOES NOT HAVE A US 510K NUMBER. HOWEVER, THIS MDR IS BEING SUBMITTED BECAUSE IT IS THE SAME AS OR SIMILAR TO A PRODUCT DISTRIBUTED WITHIN THE US. THE DEVICE IS CURRENTLY IN THE PROCESS OF BEING EVALUATED AT THE FACILITY BY A BAXTER FIELD SERVICE TECHNICIAN. A FOLLOW-UP REPORT WILL BE FILED UPON COMPLETION OF THE EVALUATION OR IF ANY ADDITIONAL DETAILS BECOME AVAILABLE.
(B)(4). EVALUATION SUMMARY: THE REPORTED CONDITION OF A COLLEAGUE INFUSION PUMP WITH A BATTERY ISSUE WAS CONFIRMED BY BAXTER PERSONNEL AS A DEPLETED BATTERY ALARM. THIS CONDITION WAS CAUSED BY DEPLETED MAIN BATTERIES. THIS CONDITION WAS RESOLVED AT THE FACILITY BY A BAXTER FIELD SERVICE TECHNICIAN BY REPLACING THE MAIN BATTERIES AND BATTERY HARNESS. ADDITIONAL INFORMATION: THIS DEVICE IS A REMEDIATED COLLEAGUE PUMP WITH A USER INTERFACE MODULE SOFTWARE VERSION OF 5.08.92. BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. THE ROOT CAUSE INVESTIGATION IS IN PROGRESS THROUGH (B)(4). SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, A FOLLOW-UP MEDWATCH WILL BE SUBMITTED.
THE FACILITY REPORTED TO BAXTER (B)(4) THAT A COLLEAGUE INFUSION PUMP EXPERIENCED A BATTERY ISSUE. IT IS UNKNOWN WHEN OR IN WHICH CARE AREA THIS EVENT OCCURRED. UPON REVIEWING THE PUMP'S EVENT HISTORY, BAXTER QUALITY ENGINEERING DISCOVERED THIS EVENT INTERRUPTED DELIVERY. THERE WAS NO PATIENT INJURY, MEDICAL INTERVENTION NECESSARY, OR ADVERSE REACTION IN ASSOCIATION WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE. THE USER INTERFACE MODULE SOFTWARE VERSION OF THIS DEVICE IS CURRENTLY UNKNOWN.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | COLLEAGUE SINGLE CHANNEL VOLUMETRIC PUMP CE FRENCH | PUMP, INFUSION | FRN | BAXTER HEALTHCARE - SINGAPORE |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |