HALF DAY INFUSOR, 5ML/HR 12PK
Report
- Report Number
- 6000001-2010-05045
- Event Type
- Malfunction
- Date Received
- November 22, 2010
- Date of Event
- October 1, 2010
- Report Date
- October 21, 2010
- Manufacturer
- BAXTER HEALTHCARE - IRVINE
- Product Code
- MEB
- PMA / PMN Number
- K982012
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UK
- Reporter Occupation
- OTHER
Narratives
(B)(4). DEVICE EVALUATION: ONE SAMPLE WAS RECEIVED BY BAXTER FOR EVALUATION. THE REPORTED CONDITION OF "RUPTURED RESERVOIR" WAS CONFIRMED. THIS IS A SINGLE USE DEVICE AND WILL NOT BE REPAIRED. BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. THE ROOT CAUSE WILL BE IDENTIFIED/ADDRESSED THROUGH THE CAPA INVESTIGATION, (B)(4). A BATCH REVIEW WAS CONDUCTED AND NO ISSUES WERE FOUND RELATED TO THE REPORTED CONDITION DURING THE MANUFACTURE OF THE LOT.
(B)(4).ADDITIONAL NARRATIVE: THE DEVICE IS AVAILABLE FOR EVALUATION PER THE CUSTOMER; HOWEVER, THE DEVICE HAS NOT YET BEEN RECEIVED BY BAXTER. SHOULD THE DEVICE AND/OR ANY ADDITIONAL INFORMATION BECOME AVAILABLE, A FOLLOW-UP REPORT WILL BE SUBMITTED. A BATCH REVIEW WAS CONDUCTED AND NO ISSUES WERE FOUND RELATED TO THE REPORTED CONDITION DURING THE MANUFACTURE OF THE LOT.
IT WAS REPORTED TO BAXTER (B)(4) THAT THE RESERVOIR OF ONE (1) HALF DAY INFUSOR HAD RUPTURED DURING FILLING WITH DESFERRIOXAMINE. NO PATIENT INJURY OR MEDICAL INTERVENTION HAS BEEN REPORTED BY THE CUSTOMER; NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | HALF DAY INFUSOR, 5ML/HR 12PK | PUMP, INFUSION, ELASTOMERIC | MEB | BAXTER HEALTHCARE - IRVINE | 10G003 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |