FDA Adverse Event Malfunction Summary report: N

AUTOMATED PD SET W/CASSETTE4 PRONG

MDR report key: 1904785 · Received November 22, 2010

Report

Report Number
1423500-2010-05964
Event Type
Malfunction
Date Received
November 22, 2010
Date of Event
October 28, 2010
Report Date
October 28, 2010
Manufacturer
BAXTER HEALTHCARE - MOUNTAIN HOME
Product Code
FKX
PMA / PMN Number
K923065
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). DURING INVESTIGATION BY BAXTER, THIS INCIDENT WAS DETERMINED TO BE CAUSED BY USE/USER ERROR. THERE WAS NO ALLEGATION OF A PRODUCT MALFUNCTION; THEREFORE, A BATCH REVIEW AND SAMPLE EVALUATION WILL NOT BE CONDUCTED. SHOULD ANY ADDITIONAL INFORMATION BECOME AVAILABLE, A FOLLOW-UP REPORT WILL BE SUBMITTED.

Additional Manufacturer Narrative · 1

(B)(4). AN ENGINEERING QUALITY REVIEW WAS COMPLETED FOR THIS REPORT OF A PATIENT DISCONNECTING THE PATIENT LINE TO ADD A NEW PATIENT LINE EXTENSION WHICH RESULTED IN A CHECK PATIENT LINE ALARM. THE REPORTED CONDITION WAS NOT CONFIRMED DUE TO LACK OF PRODUCT SAMPLE. THE LOT NUMBER WAS UNKNOWN; THEREFORE, A BATCH REVIEW WAS NOT PERFORMED. BASED ON THE INFORMATION OBTAINED FROM BAXTER'S INVESTIGATION, THE ROOT CAUSE OF THE INCIDENT WAS USE ERROR. THE LABELING REVIEW FOUND THE PATIENT AT HOME GUIDE TO BE ADEQUATE FOR THE USE/USER ERROR IDENTIFIED IN THIS INCIDENT. BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. BAXTER WILL CONTINUE TO MONITOR SIMILAR REPORTS TO DETERMINE IF FURTHER ACTIONS ARE REQUIRED.

Description of Event or Problem · 1

DURING TROUBLESHOOTING FOR A RELATED COMPLAINT FILE, THE HOME PATIENT (HP) REPORTED TO GLOBAL TECHNICAL SERVICES THAT SHE DISCONNECTED THE PATIENT LINE AND ADDED A NEW PATIENT LINE EXTENSION. THE TSR ADVISED THE HP SHE SHOULD NOT DISCONNECT AND ADD A NEW LINE. THE TSR ADVISED TO END THERAPY AND INFORM THE REGISTERED NURSE (RN) OF EVENT. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION INDICATED AT THE TIME OF THE INITIAL REPORT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 AUTOMATED PD SET W/CASSETTE4 PRONG SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - MOUNTAIN HOME

Patients

Seq Age Sex Outcome Treatment
1