PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM
Report
- Report Number
- 2953144-2010-02896
- Event Type
- Injury
- Date Received
- November 22, 2010
- Date of Event
- October 8, 2010
- Report Date
- October 26, 2010
- Manufacturer
- AV-REDWOOD CITY
- Product Code
- MGB
- PMA / PMN Number
- P960043
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IA, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
(B)(4):DURING PROCESSING OF THIS COMPLAINT, ATTEMPTS WERE MADE TO OBTAIN COMPLETE EVENT, PATIENT AND DEVICE INFORMATION.THE DEVICE WAS RECEIVED. INVESTIGATION IS NOT YET COMPLETE.
(B)(4). EVALUATION OF THE RETURNED DEVICE FOUND THE PLUNGER WITH THE ANTERIOR NEEDLE TIP, LINK, AND CUFFS WERE NOT RETURNED. THE SHEATH, GUIDE, FOOT, AND LEVER WERE FOUND UNDAMAGED. THE SUTURE WAS RETURNED LOADED IN THE DEVICE WITH THE KNOT UNRAVELED AND THE POSTERIOR NEEDLE TIP LOOSE. THESE FINDINGS ARE CONSISTENT WITH AND CONFIRM THE REPORTED NEEDLE TO CUFF MISS EXPERIENCE. THE POSTERIOR AND ANTERIOR FOOT WERE EXAMINED FOR NEEDLE STRIKE MARKS AND THERE WERE NONE DETECTED, INDICATING THE NEEDLES WERE DEFLECTED OUTSIDE OF THE FOOT POCKET. THE SUTURE WAS PULLED FROM THE DEVICE WITH NO SIGNIFICANT RESISTANCE DETECTED INDICATING THE SUTURE DRAG WAS NOT A CONTRIBUTING FACTOR IN THIS EVENT. A PROXY PLUNGER WAS INSERTED TO TEST THE NEEDLE TRAJECTORY AND THE RESULT WAS ACCEPTABLE. BASED ON THE INVESTIGATION FINDINGS, THE MOST PROBABLE ROOT CAUSE FOR THE POSTERIOR CUFF MISS AND SUBSEQUENT FAILURE TO ACHIEVE HEMOSTASIS IS DUE TO NEEDLE DEFLECTION DURING PLUNGER DEPLOYMENT POSSIBLY CAUSED BY INTERACTION WITH ANATOMY OR A FAILURE TO MAINTAIN A STABLE POSITION OF THE DEVICE WITH RESPECT TO THE TISSUE TRACT DURING NEEDLE DEPLOYMENT. NO MANUFACTURING OR QUALITY ISSUES WERE DETECTED. REVIEW OF THE DEVICE HISTORY RECORD FOR THIS LOT DID NOT PRODUCE ANY FINDINGS RELEVANT TO THIS REPORT.
IT WAS REPORTED THAT A PHYSICIAN TRAINED IN THE USE OF THE PROGLIDE DEVICE ATTEMPTED ARTERIOTOMY CLOSURE OF THE COMMON FEMORAL ARTERY AFTER AN UNSPECIFIED PROCEDURE. REPORTEDLY, A CUFF MISS OCCURRED AND A SECOND PROGLIDE WAS USED SUCCESSFULLY TO ACHIEVE HEMOSTASIS. ALTHOUGH REQUESTED, NO ADDITIONAL INFORMATION WAS PROVIDED.
IT WAS REPORTED BY THE DISTRIBUTOR THAT THE SIDE RAILS DO NOT CLOSE PROPERLY. NO PT INVOLVEMENT OR ADVERSE CONSEQUENCES WERE REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM | SUTURE MEDIATED CLOSURE | MGB | AV-REDWOOD CITY | 930106H |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |