PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM
Report
- Report Number
- 2953144-2010-02872
- Event Type
- Injury
- Date Received
- November 22, 2010
- Date of Event
- October 11, 2010
- Report Date
- October 26, 2010
- Manufacturer
- AV-REDWOOD CITY
- Product Code
- MGB
- PMA / PMN Number
- P960043
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IA, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
(B)(4). THE DEVICE WAS RECEIVED. INVESTIGATION IS NOT YET COMPLETE.(B)(4):DURING PROCESSING OF THIS COMPLAINT, ATTEMPTS WERE MADE TO OBTAIN COMPLETE EVENT, PATIENT AND DEVICE INFORMATION.
(B)(4). EVALUATION SUMMARY: THE WHOLE SUTURE WAS RETURNED UNDAMAGED AND PARTIALLY EXPOSED. THE PLUNGER, THE ANTERIOR NEEDLE AND ANTERIOR CUFF WERE NOT RETURNED. EVALUATION OF THE RETURNED DEVICE FOUND THAT THE POSTERIOR CUFF MISS OCCURRED AS EVIDENCED BY THE POSTERIOR CUFF REMAINING IN THE FOOT POCKET. THE POSTERIOR NEEDLE TIP WAS BENT AND THERE WAS A NEEDLE STRIKE MARK AT THE POSTERIOR FOOT. THIS IS INDICATIVE OF POSTERIOR NEEDLE STRIKING THE FOOT INSTEAD OF ENGAGING WITH THE POSTERIOR CUFF INSIDE THE POSTERIOR FOOT POCKET. BECAUSE THE POSTERIOR NEEDLE DID NOT ENGAGE WITH THE POSTERIOR CUFF, THE CUFF WAS NOT EJECTED FROM THE FOOT. WHEN THE PLUNGER WAS REMOVED FROM THE DEVICE, THE LINK WAS HELD ON ONE END BY THE POSTERIOR CUFF IN THE FOOT POCKET WHILE BEING PULLED ON THE OTHER END BY THE WITHDRAWAL OF THE PLUNGER, RESULTING IN THE DETACHING FROM THE ANTERIOR CUFF. DURING LAB TESTING, A PROXY PLUNGER WAS INSERTED TO TEST THE NEEDLE TRAJECTORY AND PUSH MANDREL TRAVEL LENGTH AND THE RESULTS WERE ACCEPTABLE. BASED ON THE INVESTIGATION FINDINGS, THE PROBABLE ROOT CAUSE FOR THE POSTERIOR CUFF MISS AND SUBSEQUENT LINK BREAK AT THE ANTERIOR CUFF IS NEEDLE DEFLECTION DUE TO INTERACTION WITH HUMAN TISSUE OR A FAILURE TO MAINTAIN A STABLE POSITION OF THE DEVICE WITH RESPECT TO THE TISSUE TRACT. NO MANUFACTURING OR QUALITY ISSUE WAS DETECTED. REVIEW OF THE DEVICE HISTORY RECORD FOR THIS LOT DID NOT PRODUCE ANY FINDINGS RELEVANT TO THIS REPORT.
IT WAS REPORTED THAT A PHYSICIAN TRAINED IN THE USE OF THE PROGLIDE DEVICE ATTEMPTED ARTERIOTOMY CLOSURE OF THE COMMON FEMORAL ARTERY AFTER AN UNSPECIFIED PROCEDURE. REPORTEDLY, A CUFF MISS OCCURRED AND A SECOND PROGLIDE WAS USED SUCCESSFULLY TO ACHIEVE HEMOSTASIS. ALTHOUGH REQUESTED, NO ADDITIONAL INFORMATION WAS PROVIDED.
ON (B)(6) 2010, PT REPORTED THE BUTTONS ON HIS INFUSION DEVICE ARE FUNCTIONING INTERMITTENTLY. PT STATED HE STARTED NOTICING THE ISSUE A COUPLE OF WEEKS AGO. PT REPORTED THE UP AND DOWN BUTTONS ARE NOT ALWAYS RESPONDING FOR HIM. PT STATED HE NOTICED THIS ISSUE WHILE ATTEMPTING TO BOLUS. PT IS NOT ABLE TO SEE THE SCREEN VERY WELL DUE TO VISION CONCERNS. PT REPORTED THE BUTTONS POP BACK UP AFTER BEING PRESSED. NO PHYSIOLOGICAL EFFECTS WERE REPORTED. THE PT DID NOT REQUIRE ASSISTANCE FROM A HEALTHCARE PROFESSIONAL OR SECOND PARTY TO ADDRESS THE ISSUE. PRODUCT WAS REPLACED AND REQUESTED RETURN OF THE ALLEGED PRODUCT FOR EVAL.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM | SUTURE MEDIATED CLOSURE | MGB | AV-REDWOOD CITY | 930106H |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |