FDA Adverse Event Malfunction Summary report: N

MINICAP EXTEND LIFE PD TRANSFSET W/TWIST CLAMP

MDR report key: 1904522 · Received November 22, 2010

Report

Report Number
1423500-2010-06086
Event Type
Malfunction
Date Received
November 22, 2010
Date of Event
November 1, 2010
Report Date
November 1, 2010
Manufacturer
BAXTER HEALTHCARE - MOUNTAIN HOME
Product Code
KDJ
PMA / PMN Number
K882498
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
HK
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE ACTUAL SAMPLE WAS RECEIVED AND EVALUATED. THIS COMPLAINT FOR A LEAK WAS CONFIRMED IN THE LAB. THE RESULTS OF A SAMPLE EVALUATION REVEALED THAT BOTH OF THE OCCLUDER FEET WERE BROKEN. A ROOT CAUSE WAS NOT IDENTIFIED DUE TO INSUFFICIENT INFORMATION. SIMILAR REPORTS HAVE BEEN RECEIVED BY BAXTER FOR THE REPORTED PROBLEM. BAXTER WILL CONTINUE TO MONITOR SIMILAR REPORTS TO DETERMINE IF FURTHER ACTIONS ARE REQUIRED. A BATCH REVIEW WILL NOT BE PERFORMED BECAUSE THE LOT INFORMATION IS UNKNOWN.

Additional Manufacturer Narrative · 1

(B)(4). THE PRODUCT SAMPLE HAS BEEN REQUESTED BUT HAS NOT YET BEEN RECEIVED BY BAXTER FOR EVALUATION. SHOULD ANY ADDITIONAL INFORMATION BECOME AVAILABLE, A FOLLOW-UP REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

THE CUSTOMER REPORTED TO BAXTER THAT THEY FOUND TRANSFER SET LEAKAGE AFTER CLOSING IT. THIS PROBLEM WAS FOUND DURING PATIENT USE. THERE WAS NO PATIENT INJURY REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 MINICAP EXTEND LIFE PD TRANSFSET W/TWIST CLAMP SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE KDJ BAXTER HEALTHCARE - MOUNTAIN HOME

Patients

Seq Age Sex Outcome Treatment
1