FDA Adverse Event Malfunction Summary report: N

ESPRIT VENTILATOR

MDR report key: 1904458 · Received November 22, 2010

Report

Report Number
2031642-2010-00299
Event Type
Malfunction
Date Received
November 22, 2010
Date of Event
October 22, 2010
Report Date
October 26, 2010
Manufacturer
RESPIRONICS CALIFORNIA, INC.
Product Code
CBK
PMA / PMN Number
K981072
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

PRINTED CIRCUIT BOARD (PCB); SOLENOID.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THE VENTILATOR FAILED PRE-OPERATIONAL TESTING OF THE VENTILATOR DUE TO AN INHALATION AUTOZERO SOLENOID FAILURE. THERE WAS NO PATIENT INVOLVEMENT AND THEREFORE NO PATIENT HARM. IF THE SOLENOID MALFUNCTIONS AND CANNOT OPEN, THIS TASK CANNOT BE PERFORMED AND AFFECTS THE ACCURACY OF THE SYSTEM PRESSURE MEASUREMENT. FAILURE OF THE AUTOZERO SOLENOIDS WHILE IN USE COULD RESULT IN A VENT INOP CONDITION. VENT INOP, IF IN USE ON A PATIENT, WILL STOP PROVIDING VENTILATORY SUPPORT TO THE PATIENT. THE SERVICE TECHNICIAN REPLACED THE SENSOR BOARD AND THREE STATION SOLENOID TO CORRECT THE PROBLEM. EXTENDED SELF TESTING AND PERFORMANCE VERIFICATION TESTING WAS COMPLETED AND TESTS PASSED TO OPERATING SPECIFICATIONS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ESPRIT VENTILATOR VENTILATOR, CONTINUOUS CBK RESPIRONICS CALIFORNIA, INC. V1000 NA

Patients

Seq Age Sex Outcome Treatment
1