FDA Adverse Event Malfunction Summary report: N

HS HAND CONTROLLED CURVED SHEA

MDR report key: 1904445 · Received February 8, 2008

Report

Report Number
1527736-2008-00765
Event Type
Malfunction
Date Received
February 8, 2008
Date of Event
December 10, 2007
Report Date
December 19, 2007
Manufacturer
ETHICON ENDO SURGERY, INC (CINCINNATI)
Product Code
LFL
PMA / PMN Number
K002981
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KS, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: THE ANALYSIS RESULTS FOUND THAT THE DEVICE WAS RETURNED WITH THE BLADE SCRATCHED AND CRACKED. DUE TO THE DAMAGE TO THE BLADE, IT WAS CONFIRMED THAT THE DEVICE WAS NON-FUNCTIONAL. THE DEVICE WAS TESTED WITH A GENERATOR AND AN ERROR CODE 5 WAS NOTED. THE IDENTIFIED BLADE DAMAGE MAY HAVE OCCURRED FROM EXTERNAL CONTACT DURING PRE-OP OR GENERAL USE. IN ADDITION, MINOR BLADE DAMAGE MAY INCREASE IN SEVERITY DURING SUBSEQUENT ACTIVATIONS, AND MAY RESULT IN BLADE "LOCKOUT" LATER IN THE PROCEDURE. FOR THIS REASON THE FOLLOWING STATEMENTS WERE INCLUDED IN THE INSTRUCTIONS FOR USE: "AVOID ACCIDENTAL CONTACT WITH ALL METAL OR PLASTIC INSTRUMENTS OR OBJECTS WHEN THE INSTRUMENT IS ACTIVATED. CONTACT WITH STAPLES, CLIPS OR OTHER INSTRUMENTS WHILE THE INSTRUMENT IS ACTIVATED MAY RESULT IN CRACKED OR BROKEN BLADES, WHICH MAY BE IDENTIFIED BY GENERATOR SOLID TONE OR INSTRUMENT ERROR." EACH DEVICE IS VISUALLY INSPECTED AND FUNCTIONALLY TESTED PRIOR TO SHIPMENT, AND DAMAGE OF THIS MAGNITUDE WOULD HAVE BEEN DETECTED AT THIS PROCESS. COMPLAINT INFO IS TRENDED ON A REGULAR BASIS TO DETERMINE IF FURTHER INVESTIGATION IS WARRANTED. A BATCH RECORD REVIEW WAS PERFORMED AND NO ANOMALIES WERE FOUND DURING THE MANUFACTURING PROCESS.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING AN UNK PROCEDURE, AN INSTRUMENT FAILURE WAS RECEIVED. ANOTHER INSTRUMENT WAS USED TO COMPLETE THE PROCEDURE WITH NO PT CONSEQUENCE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 HS HAND CONTROLLED CURVED SHEA GEI LFL ETHICON ENDO SURGERY, INC (CINCINNATI) NA D4JE2M

Patients

Seq Age Sex Outcome Treatment
1 HANDPIECE| GENERATOR