FDA Adverse Event Injury Summary report: N

HYDRO THERMABLATOR (HTA) ENDOMETRIAL ABLATION SYSTEM

MDR report key: 1904440 · Received November 22, 2010

Report

Report Number
3005099803-2010-04743
Event Type
Injury
Date Received
November 22, 2010
Date of Event
October 29, 2010
Report Date
November 1, 2010
Manufacturer
BOSTON SCIENTIFIC - MARLBOROUGH
Product Code
MNB
PMA / PMN Number
P000040
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
MO, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE COMPLAINANT WAS UNABLE TO PROVIDE THE LOT NUMBER, THEREFORE, EXPIRATION AND MANUFACTURING DATES CANNOT BE DETERMINED. THE COMPLAINANT HAS INDICATED THAT THE PRODUCT HAS BEEN DISPOSED; THEREFORE, A FAILURE ANALYSIS OF THE COMPLAINT DEVICE COULD NOT BE COMPLETED. AT THIS TIME, WE ARE UNABLE TO DETERMINE THE RELATIONSHIP BETWEEN THE DEVICE AND THE CAUSE OF THIS EVENT. IF ANY ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL MEDWATCH WILL BE FILED.

Description of Event or Problem · 1

A HYDROTHERMABLATION PROCERVA PROCEDURE SET WAS USED DURING A HYDROTHERMABLATION (HTA) PROCEDURE. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, THE PATIENT SUDDENLY MOVED AND THEN COUGHED. AS A RESULT, FLUID WAS OBSERVED AT THE UTERUS AND THE SMALL PARTS OF THE BUTTOCK'S WERE BLISTERED. CLINICAL FOLLOW UP INFORMATION REVEALED THAT A TENACULUM WAS USED TO MAINTAIN A CERVICAL SEAL AND A TENACULUM STABILIZER WAS ALSO USED, THE BURN WAS CATEGORIZED AS "2ND DEGREE", A FLUID LOSS ALARM WAS GENERATED AT 10 CC'S, BURNS WERE BOTH INTERNAL AND EXTERNAL AND BURNS WERE TREATED WITH LIDOCAINE AND SILVADENE CREAM. THERE WERE NO FURTHER PATIENT COMPLICATIONS AND THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE "OK". AN ADDITIONAL ATTEMPT HAS BEEN MADE TO OBTAIN FOLLOW UP EVENT DETAILS WITH NO RESPONSE FROM THE CLINICIAN.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 HYDRO THERMABLATOR (HTA) ENDOMETRIAL ABLATION SYSTEM DEVICE, THERMAL ABLATION, ENDOMETRIAL MNB BOSTON SCIENTIFIC - MARLBOROUGH M006560211

Patients

Seq Age Sex Outcome Treatment
1 47 YR Required Intervention