CPS DUO

MDR report key: 1904434 · Received November 22, 2010

Report

Report Number
2017865-2010-05360
Event Type
Malfunction
Date Received
November 22, 2010
Date of Event
April 26, 2010
Manufacturer
ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
Product Code
DQY
PMA / PMN Number
NA
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED. ANALYSIS OF THE CPS-DUO UNIT FOUND THAT THE STYLET WAS BROKEN AT 11.6 CM FROM THE WHITE HUB.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING POSITIONING OF THE LEFT VENTRICULAR LEAD, THE CPS-DUO ADAPTER PIN DISCONNECTED. DUE TO THIS, THE STYLET BROKE AND STUCK INSIDE THE LEAD. THE LEAD WAS REPLACED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CPS DUO GUIDEWIRE DQY ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION DS2M007 NA

Patients

Seq Age Sex Outcome Treatment
1