FDA Adverse Event
Malfunction
Summary report: N
CPS DUO
MDR report key: 1904434
·
Received November 22, 2010
Report
- Report Number
- 2017865-2010-05360
- Event Type
- Malfunction
- Date Received
- November 22, 2010
- Date of Event
- April 26, 2010
- Manufacturer
- ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
- Product Code
- DQY
- PMA / PMN Number
- NA
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED. ANALYSIS OF THE CPS-DUO UNIT FOUND THAT THE STYLET WAS BROKEN AT 11.6 CM FROM THE WHITE HUB.
Description of Event or Problem · 1
IT WAS REPORTED THAT DURING POSITIONING OF THE LEFT VENTRICULAR LEAD, THE CPS-DUO ADAPTER PIN DISCONNECTED. DUE TO THIS, THE STYLET BROKE AND STUCK INSIDE THE LEAD. THE LEAD WAS REPLACED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CPS DUO | GUIDEWIRE | DQY | ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION | DS2M007 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |