FDA Adverse Event Malfunction Summary report: N

ULTRACISION HARMONIC SCALPEL CURVED SHEARS, SCISSOR-GRIP HANDLE - 14 CM

MDR report key: 1904401 · Received January 30, 2008

Report

Report Number
1527736-2008-00589
Event Type
Malfunction
Date Received
January 30, 2008
Date of Event
November 29, 2007
Report Date
December 21, 2007
Manufacturer
ETHICON ENDO SURGERY, INC. (CINCINNATI)
Product Code
GEI
PMA / PMN Number
K002981
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). OTHER # = D9DE7U (BATCH # FOR DEVICE B); EXP DATE = 02/2012 (DEVICE B). (DEVICE B): 3/2007. EVAL SUMMARY: DEVICE A WAS RETURNED WITH THE BLADE SCRATCHED AND CRACKED AND DEVICE B WAS RETURNED WITH THE BLADE GOUGED AND CRACKED. DUE TO THE DAMAGE TO THE BLADES, IT WAS CONFIRMED THAT THE DEVICES WERE NON-FUNCTIONAL. THE DEVICES WERE TESTED WITH A GENERATOR AND AN ERROR CODE 5 WAS NOTED. THE IDENTIFIED BLADE DAMAGE MAY HAVE OCCURRED FROM EXTERNAL CONTACT DURING PRE-OP OR GENERAL USE. IN ADDITION, MINOR BLADE DAMAGE MAY INCREASE IN SEVERITY DURING SUBSEQUENT ACTIVATIONS, AND MAY RESULT IN BLADE "LOCKOUT" LATER IN THE PROCEDURE. FOR THIS REASON, THE FOLLOWING STATEMENTS WERE INCLUDED IN THE INSTRUCTIONS FOR USE: "AVOID ACCIDENTAL CONTACT WITH ALL METAL OR PLASTIC INSTRUMENTS OR OBJECTS WHEN THE INSTRUMENT IS ACTIVATED. CONTACT WITH STAPLES, CLIPS OR OTHER INSTRUMENTS WHILE THE INSTRUMENT IS ACTIVATED MAY RESULT IN CRACKED OR BROKEN BLADES, WHICH MAY BE IDENTIFIED BY GENERATOR SOLID TONE OR INSTRUMENT ERROR." NO ASSEMBLY OR DISASSEMBLY ISSUES WERE NOTED. EACH DEVICE IS VISUALLY INSPECTED AND FUNCTIONALLY TESTED PRIOR TO SHIPMENT, AND DAMAGE OF THIS MAGNITUDE WOULD HAVE BEEN DETECTED AT THIS PROCESS. WE HAVE DOCUMENTED THE CIRCUMSTANCES AS THEY WERE REPORTED TO US. IN ADDITION, COMPLAINT INFO IS TRENDED ON A REGULAR BASIS TO DETERMINE IF FURTHER INVESTIGATION IS WARRANTED. A BATCH RECORD REVIEW WAS PERFORMED AND NO ANOMALIES WERE FOUND DURING THE MFG PROCESS.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A LOBECTOMY, THERE WAS A POOR CONNECTION. ANOTHER DEVICE WAS USED TO COMPLETE THE CASE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ULTRACISION HARMONIC SCALPEL CURVED SHEARS, SCISSOR-GRIP HANDLE - 14 CM GEI ETHICON ENDO SURGERY, INC. (CINCINNATI) NA UNK

Patients

Seq Age Sex Outcome Treatment
1 GENERATOR| HANDPIECE