HARMONIC ACE 36 CM

MDR report key: 1904393 · Received January 30, 2008

Report

Report Number
1527736-2008-00587
Event Type
Malfunction
Date Received
January 30, 2008
Report Date
November 21, 2007
Manufacturer
ETHICON ENDO SURGERY, INC. (CINCINNATI)
Product Code
LFL
PMA / PMN Number
K042777
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). SCRATCHES, CRACKED. EVAL SUMMARY: DEVICES A AND B WERE RETURNED WITH THE BLADES SCRATCHED AND CRACKED. DUE TO THE DAMAGE TO THE BLADES, IT WAS CONFIRMED THAT THE DEVICES WERE NON-FUNCTIONAL. THE DEVICES WERE TESTED WITH A GENERATOR AND AN ERROR CODE 5 WAS NOTED. THE SMOKE ISSUES COULD NOT BE EVALUATED DUE TO THE RETURNED CONDITION. THE IDENTIFIED BLADE DAMAGE MAY HAVE OCCURRED FROM EXTERNAL CONTACT DURING PRE-OP OR GENERAL USE. IN ADDITION, MINOR BLADE DAMAGE MAY INCREASE IN SEVERITY DURING SUBSEQUENT ACTIVATIONS, AND MAY RESULT IN BLADE "LOCKOUT" LATER IN THE PROCEDURE. FOR THIS REASON, THE FOLLOWING STATEMENTS WERE INCLUDED IN THE INSTRUCTIONS FOR USE: "AVOID ACCIDENTAL CONTACT WITH ALL METAL OR PLASTIC INSTRUMENTS OR OBJECTS WHEN THE INSTRUMENT IS ACTIVATED. CONTACT WITH STAPLES, CLIPS OR OTHER INSTRUMENTS WHILE THE INSTRUMENT IS ACTIVATED MAY RESULT IN CRACKED OR BROKEN BLADES, WHICH MAY BE IDENTIFIED BY GENERATOR SOLID TONE OR INSTRUMENT ERROR." EACH DEVICE IS VISUALLY INSPECTED AND FUNCTIONALLY TESTED PRIOR TO SHIPMENT, AND DAMAGE OF THIS MAGNITUDE WOULD HAVE BEEN DETECTED AT THIS PROCESS. WE HAVE DOCUMENTED THE CIRCUMSTANCES AS THEY WERE REPORTED TO US. IN ADDITION, COMPLAINT INFO IS TRENDED ON A REGULAR BASIS TO DETERMINE IF FURTHER INVESTIGATION IS WARRANTED. THE BATCH HISTORY RECORDS WERE REVIEWED WITH NO ANOMALIES NOTED DURING THE MFG PROCESS.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A LAP PROSTATECTOMY PROCEDURE, BEFORE USE WITH TISSUE, SMOKE CAME OUT OF THE DEVICE. A SAME LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. NO PT CONSEQUENCE REPORTED. NO FURTHER INFO AVAILABLE.