HARMONIC ACE 36 CM
Report
- Report Number
- 1527736-2008-00587
- Event Type
- Malfunction
- Date Received
- January 30, 2008
- Report Date
- November 21, 2007
- Manufacturer
- ETHICON ENDO SURGERY, INC. (CINCINNATI)
- Product Code
- LFL
- PMA / PMN Number
- K042777
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- OTHER
Narratives
(B)(4). SCRATCHES, CRACKED. EVAL SUMMARY: DEVICES A AND B WERE RETURNED WITH THE BLADES SCRATCHED AND CRACKED. DUE TO THE DAMAGE TO THE BLADES, IT WAS CONFIRMED THAT THE DEVICES WERE NON-FUNCTIONAL. THE DEVICES WERE TESTED WITH A GENERATOR AND AN ERROR CODE 5 WAS NOTED. THE SMOKE ISSUES COULD NOT BE EVALUATED DUE TO THE RETURNED CONDITION. THE IDENTIFIED BLADE DAMAGE MAY HAVE OCCURRED FROM EXTERNAL CONTACT DURING PRE-OP OR GENERAL USE. IN ADDITION, MINOR BLADE DAMAGE MAY INCREASE IN SEVERITY DURING SUBSEQUENT ACTIVATIONS, AND MAY RESULT IN BLADE "LOCKOUT" LATER IN THE PROCEDURE. FOR THIS REASON, THE FOLLOWING STATEMENTS WERE INCLUDED IN THE INSTRUCTIONS FOR USE: "AVOID ACCIDENTAL CONTACT WITH ALL METAL OR PLASTIC INSTRUMENTS OR OBJECTS WHEN THE INSTRUMENT IS ACTIVATED. CONTACT WITH STAPLES, CLIPS OR OTHER INSTRUMENTS WHILE THE INSTRUMENT IS ACTIVATED MAY RESULT IN CRACKED OR BROKEN BLADES, WHICH MAY BE IDENTIFIED BY GENERATOR SOLID TONE OR INSTRUMENT ERROR." EACH DEVICE IS VISUALLY INSPECTED AND FUNCTIONALLY TESTED PRIOR TO SHIPMENT, AND DAMAGE OF THIS MAGNITUDE WOULD HAVE BEEN DETECTED AT THIS PROCESS. WE HAVE DOCUMENTED THE CIRCUMSTANCES AS THEY WERE REPORTED TO US. IN ADDITION, COMPLAINT INFO IS TRENDED ON A REGULAR BASIS TO DETERMINE IF FURTHER INVESTIGATION IS WARRANTED. THE BATCH HISTORY RECORDS WERE REVIEWED WITH NO ANOMALIES NOTED DURING THE MFG PROCESS.
IT WAS REPORTED THAT DURING A LAP PROSTATECTOMY PROCEDURE, BEFORE USE WITH TISSUE, SMOKE CAME OUT OF THE DEVICE. A SAME LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. NO PT CONSEQUENCE REPORTED. NO FURTHER INFO AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | HARMONIC ACE 36 CM | LFL | ETHICON ENDO SURGERY, INC. (CINCINNATI) | NA | D4HX86 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | GENERATOR| HANDPIECE |