FDA Adverse Event Malfunction Summary report: N

EON IMPLANTABLE PULSE GENERATOR

MDR report key: 1904326 · Received October 28, 2010

Report

Report Number
1627487-2010-01727
Event Type
Malfunction
Date Received
October 28, 2010
Date of Event
July 3, 2009
Report Date
July 3, 2009
Manufacturer
ADVANCED NEUROMODULATION SYSTEMS
Product Code
LGW
PMA / PMN Number
P010032
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
BE
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

EVALUATION: METHOD - THE DEVICE HISTORY AND STERILIZATION RECORDS WERE REVIEWED. RESULTS: THE DEVICE HISTORY AND STERILIZATION RECORDS REVIEWED WERE FOUND TO MEET SPECIFICATIONS AND NO ANOMALIES WERE FOUND. IPG PASSES ALL TESTING. ANTENNA HAD BEEN CUT. CONCLUSION: THE CAUSE OF THE REPORTED COMPLAINT COULD NOT BE DETERMINED FROM THE REVIEW OF THE DHR AND STERILIZATION RECORDS. (B)(4). ANS HAS LIMITED INFORMATION RELATED TO THE PATIENT'S MEDICAL HISTORY AND IS UNABLE TO FORM AN OPINION AS TO THE RELEVANCY OF THE PATIENT'S HISTORY TO THE EVENT REPORTED. ANS DEFERS TO THE PATIENT'S PHYSICIAN REGARDING MEDICAL HISTORY.

Description of Event or Problem · 1

THE PATIENT ((B)(6)) RECEIVED AN SCS SYSTEM WHICH INCLUDED AN IPG ON (B)(6) 2009. IT WAS REPORTED THE DEVICE WAS EXPLANTED AND REPLACED ON (B)(6) 2009 AS IT COULD NOT BE CHARGED. THE EXPLANTED IPG WAS RETURNED TO THE MANUFACTURER FOR EVALUATION. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 EON IMPLANTABLE PULSE GENERATOR TOTALLY IMPLANTABLE PULSE GENERATOR LGW ADVANCED NEUROMODULATION SYSTEMS 3716 115166

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention