FDA Adverse Event Malfunction Summary report: N

LAMITRODE 44C LEAD, 60CM LENGTH

MDR report key: 1904296 · Received October 28, 2010

Report

Report Number
1627487-2010-02091
Event Type
Malfunction
Date Received
October 28, 2010
Date of Event
May 2, 2008
Report Date
May 8, 2008
Manufacturer
ADVANCED NEUROMODULATION SYSTEMS
Product Code
LGW
PMA / PMN Number
P010032
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TN, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

DEVICE 1 OF 2. EVALUATION: RESULTS: LEAD FAILS CONTINUITY TEST AND HAS BROKEN WIRES. ANS HAS LIMITED INFORMATION RELATED TO THE PATIENT'S MEDICAL HISTORY AND IS UNABLE TO FORM AN OPINION AS TO THE RELEVANCY OF THE PATIENT'S HISTORY TO THE EVENT REPORTED. ANS DEFERS TO THE PATIENT'S PHYSICIAN REGARDING MEDICAL HISTORY.

Description of Event or Problem · 1

DEVICE 1 OF 2. REFERENCE MFR REPORT # 1627487-2010-02092. THE PATIENT RECEIVED HER SCS SYSTEM CONSISTING OF TWO LEADS AND AN IPG ON AN UNKNOWN DATE. THE LEADS WERE IMPLANTED IN THE OCCIPITAL AREA (OFF-LABEL PLACEMENT). IT WAS REPORTED THAT THE PATIENT STATED THE SYSTEM NEVER WORKED FOR HER AND SHE REQUESTED THE SYSTEM BE EXPLANTED. THE SYSTEM WAS EXPLANTED ON (B)(6) 2008. THE EXPLANTED DEVICES WERE RETURNED TO ANS FOR EVALUATION. NO FURTHER INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 LAMITRODE 44C LEAD, 60CM LENGTH SPINAL CORD STIMULATION LEAD LGW ADVANCED NEUROMODULATION SYSTEMS 3245

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention