LAMITRODE 44C LEAD, 60CM LENGTH
Report
- Report Number
- 1627487-2010-02091
- Event Type
- Malfunction
- Date Received
- October 28, 2010
- Date of Event
- May 2, 2008
- Report Date
- May 8, 2008
- Manufacturer
- ADVANCED NEUROMODULATION SYSTEMS
- Product Code
- LGW
- PMA / PMN Number
- P010032
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TN, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
DEVICE 1 OF 2. EVALUATION: RESULTS: LEAD FAILS CONTINUITY TEST AND HAS BROKEN WIRES. ANS HAS LIMITED INFORMATION RELATED TO THE PATIENT'S MEDICAL HISTORY AND IS UNABLE TO FORM AN OPINION AS TO THE RELEVANCY OF THE PATIENT'S HISTORY TO THE EVENT REPORTED. ANS DEFERS TO THE PATIENT'S PHYSICIAN REGARDING MEDICAL HISTORY.
DEVICE 1 OF 2. REFERENCE MFR REPORT # 1627487-2010-02092. THE PATIENT RECEIVED HER SCS SYSTEM CONSISTING OF TWO LEADS AND AN IPG ON AN UNKNOWN DATE. THE LEADS WERE IMPLANTED IN THE OCCIPITAL AREA (OFF-LABEL PLACEMENT). IT WAS REPORTED THAT THE PATIENT STATED THE SYSTEM NEVER WORKED FOR HER AND SHE REQUESTED THE SYSTEM BE EXPLANTED. THE SYSTEM WAS EXPLANTED ON (B)(6) 2008. THE EXPLANTED DEVICES WERE RETURNED TO ANS FOR EVALUATION. NO FURTHER INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | LAMITRODE 44C LEAD, 60CM LENGTH | SPINAL CORD STIMULATION LEAD | LGW | ADVANCED NEUROMODULATION SYSTEMS | 3245 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |