Summary report: N
ANIMAS INSULIN INFUSION PUMP
MDR report key: 1904160
·
Received November 15, 2010
Report
- Report Number
- 2531779-2010-02283
- Event Type
- Malfunction
- Date Received
- November 15, 2010
- Date of Event
- October 15, 2010
- Report Date
- October 15, 2010
- Manufacturer
- ANIMAS CORP.
- Product Code
- LZG
- PMA / PMN Number
- K042873
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- DE, US
- Reporter Occupation
- UNKNOWN
Narratives
Additional Manufacturer Narrative · 1
THE PUMP WAS RETURNED TO ANIMAS AND THE KEYPAD APPEARED TO BE INTACT WITH NO LIFTING OR PEELING. THE "DOWN" KEYPAD BUTTON WAS INTERMITTENTLY RESPONDING AND REQUIRED EXCESSIVE FORCE TO ACTIVATE. THE "UP" AND "OK" KEYPAD BUTTONS WERE UNRESPONSIVE. THE KEYPAD WAS REMOVED AND ALL OF THE KEY CONTACTS WERE INVERTED AND DID NOT ALWAYS SPRING BACK. THERE WAS ALSO EVIDENCE OF KEYPAD ADHESIVE FOUND UNDER ALL KEY CONTACTS.
Description of Event or Problem · 1
THE PT'S MOTHER REPORTED THAT THE BUTTONS HAVE TO BE PRESSED MULTIPLE TIMES TO ELICIT A RESPONSE FROM THE KEYPAD. THE BUTTONS FEEL NORMAL WHEN PRESSED AND THE REPORTER DENIES DAMAGE TO THE KEYPAD OR EXPOSURE TO WATER.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ANIMAS INSULIN INFUSION PUMP | INSULIN INFUSION PUMP | LZG | ANIMAS CORP. | IR 1250 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 16 YR |