FDA Adverse Event
Malfunction
Summary report: N
AFFINITY BED
MDR report key: 1904083
·
Received November 12, 2010
Report
- Report Number
- 1824206-2010-10979
- Event Type
- Malfunction
- Date Received
- November 12, 2010
- Date of Event
- October 18, 2010
- Report Date
- October 18, 2010
- Manufacturer
- HILL-ROM, INC.
- Product Code
- HDD
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WA, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE TECH FOUND THE CPR CABLE WAS TOO TIGHT WHICH CAUSED THE HEAD SECTION TO DRIFT DOWN WHEN RAISED ALL THE WAY UP. THE TECH ADJUSTED THE CPR HANDLE CABLE TO REPAIR THE BED.
Description of Event or Problem · 1
INFO RECEIVED INDICATES THE HEAD SECTION OF THE BED IS DRIFTING DOWN.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | AFFINITY BED | BIRTHING BED | HDD | HILL-ROM, INC. | 3700 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |