FDA Adverse Event Malfunction Summary report: N

AFFINITY BED

MDR report key: 1904083 · Received November 12, 2010

Report

Report Number
1824206-2010-10979
Event Type
Malfunction
Date Received
November 12, 2010
Date of Event
October 18, 2010
Report Date
October 18, 2010
Manufacturer
HILL-ROM, INC.
Product Code
HDD
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE TECH FOUND THE CPR CABLE WAS TOO TIGHT WHICH CAUSED THE HEAD SECTION TO DRIFT DOWN WHEN RAISED ALL THE WAY UP. THE TECH ADJUSTED THE CPR HANDLE CABLE TO REPAIR THE BED.

Description of Event or Problem · 1

INFO RECEIVED INDICATES THE HEAD SECTION OF THE BED IS DRIFTING DOWN.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 AFFINITY BED BIRTHING BED HDD HILL-ROM, INC. 3700

Patients

Seq Age Sex Outcome Treatment
1