FDA Adverse Event Malfunction Summary report: N

TOTALCARE SPO2RT

MDR report key: 1904054 · Received November 12, 2010

Report

Report Number
1824206-2010-10969
Event Type
Malfunction
Date Received
November 12, 2010
Date of Event
October 18, 2010
Report Date
October 18, 2010
Manufacturer
HILL-ROM, INC.
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

REPAIRS HAVE NOT BEEN COMPLETED.

Description of Event or Problem · 1

INFO RECEIVED INDICATES THE HEAD OF THE BED IS DRIFTING SLOWLY AND STAYING IN TRENDELENBURG.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TOTALCARE SPO2RT AC-POWERED ADJUSTABLE HOSPITAL BED FNL HILL-ROM, INC. 1900

Patients

Seq Age Sex Outcome Treatment
1