FDA Adverse Event Malfunction Summary report: N

UNICEL® DXC 800 PRO SYNCHRON® CHEMISTRY ANALYZER

MDR report key: 1904024 · Received November 21, 2010

Report

Report Number
2050012-2010-01357
Event Type
Malfunction
Date Received
November 21, 2010
Date of Event
October 24, 2010
Report Date
November 19, 2010
Manufacturer
BECKMAN COULTER INC.
Product Code
JJE
PMA / PMN Number
K042291
Removal / Correction Number
N/A
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A CLOT WAS LODGED IN THE EIC CAUSING IT TO OVERFLOW. THE BCI HOTLINE GUIDED THE CUSTOMER THROUGH DISASSEMBLY AND CLEANING OF THE CUP, WHICH RESOLVED THE ISSUE.

Description of Event or Problem · 1

A CUSTOMER CONTACTED BECKMAN COULTER INC (BCI) STATING THAT THE ELECTROLYTE INJECTION CUP (EIC) WAS OVERFLOWING. NO INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 UNICEL® DXC 800 PRO SYNCHRON® CHEMISTRY ANALYZER CLINICAL CHEMISTRY ANALYZER JJE BECKMAN COULTER INC. DXC 800 PRO N/A

Patients

Seq Age Sex Outcome Treatment
1