FDA Adverse Event
Malfunction
Summary report: N
UNICEL® DXC 800 PRO SYNCHRON® CHEMISTRY ANALYZER
MDR report key: 1904024
·
Received November 21, 2010
Report
- Report Number
- 2050012-2010-01357
- Event Type
- Malfunction
- Date Received
- November 21, 2010
- Date of Event
- October 24, 2010
- Report Date
- November 19, 2010
- Manufacturer
- BECKMAN COULTER INC.
- Product Code
- JJE
- PMA / PMN Number
- K042291
- Removal / Correction Number
- N/A
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NC, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A CLOT WAS LODGED IN THE EIC CAUSING IT TO OVERFLOW. THE BCI HOTLINE GUIDED THE CUSTOMER THROUGH DISASSEMBLY AND CLEANING OF THE CUP, WHICH RESOLVED THE ISSUE.
Description of Event or Problem · 1
A CUSTOMER CONTACTED BECKMAN COULTER INC (BCI) STATING THAT THE ELECTROLYTE INJECTION CUP (EIC) WAS OVERFLOWING. NO INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | UNICEL® DXC 800 PRO SYNCHRON® CHEMISTRY ANALYZER | CLINICAL CHEMISTRY ANALYZER | JJE | BECKMAN COULTER INC. | DXC 800 PRO | N/A |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |