FDA Adverse Event Malfunction Summary report: N

XPS® BUR - UNKNOWN

MDR report key: 19039553 · Received April 4, 2024

Report

Report Number
1045254-2024-00483
Event Type
Malfunction
Date Received
April 4, 2024
Date of Event
March 10, 2024
Report Date
June 25, 2024
Manufacturer
MEDTRONIC XOMED INC.
Product Code
EQJ
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.

Additional Manufacturer Narrative · 0

H3: H3: ANALYSIS RESULTS WERE NOT AVAILABLE AS OF THE DATE OF THIS REPORT. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN ANALYSIS IS COMPLETE. H6: PREVIOUSLY ADDED IMDRF ANNEX CODES B17, C20, D14 ARE NO LONGER VALID. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.

Additional Manufacturer Narrative · 0

H3: ANALYSIS FOUND THAT THERE WAS A BROKEN BUR STUCK INSIDE THE BUR PIPE OF HANDPIECE. VISUALLY, ONLY A PORTION OF THE BROKEN BUR WAS RETURNED (SHANK). THE SHANK HAD TRACES OF DISCOLORATION. THERE WERE ALSO TRACES OF STRIATIONS NOTED AT THE DISTAL END OF THE SHANK. THE OVERALL LENGTH OF THE BUR SHALL MEASURE 3.850 +0.015/-0.010 INCHES AND MEASURED 0.190 INCHES FROM THE PROXIMAL END OF THE SHANK TO THE BROKEN POINT. FUNCTIONAL TESTING COULD NOT BE PERFORMED DUE TO THE BROKEN STATE OF THE DEVICE. IN THE RETURNED CONDITION, THERE WAS AN OUT OF SPECIFICATION CONDITION THAT WAS RELATED TO THE COMPLAINT (DUE TO PHYSICAL DAMAGE). H6: PREVIOUSLY ADDED IMDRF CODES B21, C21, D16 AND G04041 ARE NO LONGER VALID. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.

Description of Event or Problem · 0

IT WAS REPORTED THAT SKEETER DRILL TIP SNAPPED OFF INSIDE DRILL. THERE WAS NO PATIENT INVOLVEMENT.

Description of Event or Problem · 0

ADDITIONAL INFORMATION RECEIVED STATING THAT AFTER THE BUR SNAP, THE USER WAS NOT ABLE TO RETRIEVE IT FROM THE DRILL.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2334023 XPS® BUR - UNKNOWN BUR, EAR, NOSE AND THROAT EQJ MEDTRONIC XOMED INC. XOM UNK DRILLBUR

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown