FDA Adverse Event Malfunction Summary report: N

FLO-GARD 6201 VOLUMETRIC INFUSION PUMP

MDR report key: 1903921 · Received November 20, 2010

Report

Report Number
6000001-2010-05124
Event Type
Malfunction
Date Received
November 20, 2010
Date of Event
October 1, 2010
Report Date
October 20, 2010
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FRN
PMA / PMN Number
K915522
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: THE REPORTED CONDITION OF A FLO-GARD INFUSION PUMP WITH A FALSE OCCLUSION ALARM WAS CONFIRMED DURING PRODUCT EVALUATION. THIS CONDITION WAS CAUSED BY THE PUMP'S DOOR ASSEMBLY BEING CRACKED IN ITS LATCH AND HANDLE AREAS. THE DOOR ASSEMBLY WAS REPLACED TO CORRECT THE REPORTED CONDITION. A SERVICE HISTORY REVIEW REVEALED NO PREVIOUS SERVICE EVENTS WERE RELATED TO THE REPORTED CONDITION. SHOULD ADDITIONAL INFORMATION BE RECEIVED, A FOLLOW-UP MEDWATCH WILL BE SUBMITTED.

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS RETURNED TO BAXTER AND IS CURRENTLY IN THE PROCESS OF BEING EVALUATED. A FOLLOW-UP REPORT WILL BE FILED UPON COMPLETION OF THE EVALUATION OR IF ANY ADDITIONAL DETAILS BECOME AVAILABLE.

Additional Manufacturer Narrative · 1

(B)(4). BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. BAXTER WILL CONTINUE TO MONITOR SIMILAR REPORTS TO DETERMINE IF FURTHER ACTIONS ARE REQUIRED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, A FOLLOW-UP MEDWATCH WILL BE SUBMITTED.

Description of Event or Problem · 1

THROUGH PAPERWORK RECEIVED WITH THE PUMP ON (B)(6) 2010, THE FACILITY REPORTED A FLO-GARD INFUSION PUMP WITH A CONSTANT OCCLUSION ALARM. IT IS UNKNOWN WHEN THIS EVENT OCCURRED. THERE IS NO REPORT OF PATIENT INJURY, MEDICAL INTERVENTION, OR ADVERSE REACTION IN ASSOCIATION WITH THIS EVENT. DURING PRODUCT EVALUATION, A CONSTANT OCCLUSION ALARM WAS CONFIRMED TO BE CAUSED BY THE DOOR ASSEMBLY BEING BROKEN IN THE LATCH AREA, INDICATING THE OCCLUSION ALARM WAS FALSE. NO ADDITIONAL INFORMATION IS AVAILABLE.