FDA Adverse Event Malfunction Summary report: N

UNICEL® DXC 600 PRO SYNCHRON® CHEMISTRY ANALYZER

MDR report key: 1903898 · Received November 20, 2010

Report

Report Number
2050012-2010-01227
Event Type
Malfunction
Date Received
November 20, 2010
Date of Event
October 24, 2010
Report Date
October 25, 2010
Manufacturer
BECKMAN COULTER INC.
Product Code
JGS
PMA / PMN Number
K042291
Removal / Correction Number
N/A
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NM, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

QC HAS BEEN INCONSISTENT. PER HOTLINE REQUEST, THE CUSTOMER FLUSHED AND RECALIBRATED THE MODULAR CHEMISTRY (MC) PROBE AND RAN QC WHICH WAS NOT WITHIN ACCEPTABLE RANGE. A BCI FIELD SERVICE ENGINEER (FSE) REPLACED THE MC SAMPLE: PROBE, WASH COLLAR, SYRINGE, SYRINGE VALVE. FSE REMOVED AND INSPECTED THE FLOW CELL AND DISCOVERED FIBROUS MATERIAL. FSE PERFORMED BIOBURDEN TESTING AND FOUND BACTERIAL GROWTH IN THE INTERNAL MC SIDE OF THE REAGENT AND TUBING. FSE DECONTAMINATED THE UNIT AND REPLACED ALL ISE TUBING AND CARBON BRIDGE. PART REPLACEMENT AND SYSTEM DECONTAMINATION RESOLVED THE ISSUE.

Description of Event or Problem · 1

A CUSTOMER CONTACTED BECKMAN COULTER INC. (BCI) IN REGARDS TO ERRONEOUS HIGH SODIUM (NA) RESULT GENERATED BY UNICEL DXC 600 PRO SYNCHRON CHEMISTRY ANALYZER. THE HIGH RESULT (165 MMOL/L) TRIGGERED A CRITICAL RERUN THAT RESULTED IN LOWER VALUE (141 MMOL/L) BEING OBTAINED. THE SAMPLE WAS NOT REPORTED OUT OF THE LABORATORY. PATIENT TREATMENT WAS NOT IMPACTED BY THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 UNICEL® DXC 600 PRO SYNCHRON® CHEMISTRY ANALYZER CLINICAL CHEMISTRY ANALYZER, JGS BECKMAN COULTER INC. DXC 600 PRO N/A

Patients

Seq Age Sex Outcome Treatment
1