FDA Adverse Event Malfunction Summary report: N

PRESSUREGUARD EASY AIR

MDR report key: 1903574 · Received November 12, 2010

Report

Report Number
1041130-2010-00010
Event Type
Malfunction
Date Received
November 12, 2010
Date of Event
August 24, 2010
Report Date
November 12, 2010
Manufacturer
SPAN-AMERICA MEDICAL SYSTEM, INC.
Product Code
FNM
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OR, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE PRODUCT WAS NOT RETURNED. F/U ATTEMPTS WERE MADE, BUT STILL HAVE NOT RECEIVED PRODUCT. THIS PROBLEM WAS IDENTIFIED IN HOUSE ON (B)(4) 2009 AND A CAR WAS ISSUED TO OUR MFR ON THAT DATE. WE DID SORT INTERNALLY FOR THIS SYMPTOM DURING THAT TIME. ROOT CAUSE WAS IDENTIFIED AND CORRECTIVE ACTION WERE TAKEN ON (B)(4) 2009. THE SPACER FABRIC MATERIAL DIAMETER WAS TOO LARGE AND INTERFERING WITH THE WELDING PROCESS, CAUSING POOR WELDS. THIS IS A "SPOT" PROBLEM AND NOT A CATASTROPHIC ENTIRE COVER TOP ISSUE. THIS PARTICULAR COVER WAS PART OF THE PRODUCTION RUN PRIOR TO TAKING CORRECTIVE ACTION.

Description of Event or Problem · 1

WE RECEIVED A PHONE CALL STATING THAT THEY HAD 2 WELD FAILURES ON A PRESSUREGUARD EASY AIR COVER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PRESSUREGUARD EASY AIR ALTERNATING PRESSURE MATTRESS FNM SPAN-AMERICA MEDICAL SYSTEM, INC. NA

Patients

Seq Age Sex Outcome Treatment
1