FDA Adverse Event
Malfunction
Summary report: N
PRESSUREGUARD EASY AIR
MDR report key: 1903574
·
Received November 12, 2010
Report
- Report Number
- 1041130-2010-00010
- Event Type
- Malfunction
- Date Received
- November 12, 2010
- Date of Event
- August 24, 2010
- Report Date
- November 12, 2010
- Manufacturer
- SPAN-AMERICA MEDICAL SYSTEM, INC.
- Product Code
- FNM
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OR, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE PRODUCT WAS NOT RETURNED. F/U ATTEMPTS WERE MADE, BUT STILL HAVE NOT RECEIVED PRODUCT. THIS PROBLEM WAS IDENTIFIED IN HOUSE ON (B)(4) 2009 AND A CAR WAS ISSUED TO OUR MFR ON THAT DATE. WE DID SORT INTERNALLY FOR THIS SYMPTOM DURING THAT TIME. ROOT CAUSE WAS IDENTIFIED AND CORRECTIVE ACTION WERE TAKEN ON (B)(4) 2009. THE SPACER FABRIC MATERIAL DIAMETER WAS TOO LARGE AND INTERFERING WITH THE WELDING PROCESS, CAUSING POOR WELDS. THIS IS A "SPOT" PROBLEM AND NOT A CATASTROPHIC ENTIRE COVER TOP ISSUE. THIS PARTICULAR COVER WAS PART OF THE PRODUCTION RUN PRIOR TO TAKING CORRECTIVE ACTION.
Description of Event or Problem · 1
WE RECEIVED A PHONE CALL STATING THAT THEY HAD 2 WELD FAILURES ON A PRESSUREGUARD EASY AIR COVER.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PRESSUREGUARD EASY AIR | ALTERNATING PRESSURE MATTRESS | FNM | SPAN-AMERICA MEDICAL SYSTEM, INC. | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |