FDA Adverse Event Injury Summary report: N

ASR ACETABULAR CUPS 54

MDR report key: 1903298 · Received November 16, 2010

Report

Report Number
1818910-2010-08637
Event Type
Injury
Date Received
November 16, 2010
Date of Event
April 1, 2007
Report Date
June 21, 2013
Manufacturer
DEPUY INTERNATIONAL, LTD.
Product Code
KWA
PMA / PMN Number
NA
Removal / Correction Number
Z-1749/1816-2011
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

DEPUY STILL CONSIDERS THIS INVESTIGATION CLOSED.

Additional Manufacturer Narrative · 1

DEPUY STILL CONSIDERS THIS INVESTIGATION CLOSED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THE INVESTIGATION WILL BE RE-OPENED.

Additional Manufacturer Narrative · 1

THE DEVICE ASSOCIATED WITH THIS REPORT WAS NOT RETURNED. A COMPLAINT DATABASE SEARCH AND/OR A REVIEW OF DEVICE HISTORY RECORDS WERE NOT POSSIBLE AS THE NECESSARY PRODUCT/LOT CODE COMBINATION WAS NOT PROVIDED. THE INVESTIGATION COULD NOT DRAW ANY CONCLUSIONS REGARDING THE REPORTED EVENT. ALTHOUGH THE ROOT CAUSE CANNOT BE DETERMINED, IT IS KNOWN THAT THE ASR PLATFORM WAS VOLUNTARILY RECALLED IN AUGUST OF 2010, FOLLOWING AN HHE (HEALTH HAZARD EVALUATION). FURTHER ANALYSIS WILL BE DOCUMENTED ON WWCAPA (B)(4) . BASED ON THE INABILITY TO DETERMINE A ROOT CAUSE, THE NEED FOR FURTHER CORRECTIVE ACTION IS NOT INDICATED AT THIS TIME. DEPUY CONSIDERS THE INVESTIGATION CLOSED AT THIS TIME. SHOULD THE PRODUCT AND/OR ADDITIONAL INFO BE RECEIVED, THE INVESTIGATION WILL BE RE-OPENED.

Additional Manufacturer Narrative · 1

(B)(4). THE INVESTIGATION HAS BEEN REOPENED DUE TO RECEIVING REVISION NOTIFICATION. DEPUY WILL NOTIFY THE FDA WHEN THE INVESTIGATION IS COMPLETE.

Additional Manufacturer Narrative · 1

THE ASR PLATFORM WAS VOLUNTARILY RECALLED FROM THE MARKET IN AUGUST 2010, AND THE ASR PRODUCT CODES ARE NOW CONSIDERED INACTIVE. FURTHER INVESTIGATION OF THIS INDIVIDUAL INCIDENT WILL NOT BE UNDERTAKEN, AS THERE IS AN ONGOING INVESTIGATION REGARDING THE ROOT CAUSE(S) AND/OR CORRECTIVE ACTIONS. REF. (B)(4). DEPUY CONSIDERS THE INVESTIGATION CLOSED AT THIS TIME. SHOULD THE PRODUCT AND/OR ADDITIONAL INFORMATION BE RECEIVED, THE INVESTIGATION WILL BE RE-OPENED.

Description of Event or Problem · 1

**UPDATE (B)(6) 2013 LITIGATION PAPERS RECEIVED ALLEGING THAT THE PATIENT SUFFERS FROM SQUEAKING, GRINDING,AND POPPING. ALSO ALLEGED IS FLUID COLLECTION, DISCOMFORT, AND LIMITED MOBILITY.

Description of Event or Problem · 1

PT CONTACTED (B)(6) TO INITIATE CLAIM. PT REPORTED REVISION SURGERY. REASON FOR REVISION IS UNK. NO FURTHER INFO IS AVAILABLE AT THIS TIME.

Description of Event or Problem · 1

PATIENT CONTACTED (B)(4) TO INITIATE CLAIM. PATIENT REPORTED REVISION SURGERY. REASON FOR REVISION IS UNKNOWN. NO FURTHER INFORMATION IS AVAILABLE AT THIS TIME. ***UPDATE: 01/11/2012 - THE SALES REP HAS REPORTED THE REVISION SURGERY. PATIENT WAS REVISED TO ADDRESS ACETABULAR CUP LOOSENING, PAIN AND OSTEOLYSIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ASR ACETABULAR CUPS 54 ACETABULAR CUP KWA DEPUY INTERNATIONAL, LTD. NA 2347659

Patients

Seq Age Sex Outcome Treatment
1 27 YR Required Intervention