FDA Adverse Event Malfunction Summary report: N

LIFEPAK 1000 DEFIBRILLATOR

MDR report key: 1903215 · Received November 12, 2010

Report

Report Number
3015876-2010-01247
Event Type
Malfunction
Date Received
November 12, 2010
Date of Event
October 15, 2010
Report Date
October 15, 2010
Manufacturer
PHYSIO-CONTROL, INC.
Product Code
MKJ
PMA / PMN Number
K042404
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

(B)(4): PHYSIO-CONTROL EVALUATED THE DEVICE AND VERIFIED THE REPORTED FAILURE. PHYSIO DETERMINED THE CAUSE OF THE FAILURE TO BE A FAILED INDUCTOR, DESIGNATOR, L1, FROM THE ANALOG PCB ASSEMBLY. A REPLACEMENT DEVICE WAS PROVIDED TO THE CUSTOMER.

Description of Event or Problem · 1

DURING A ROUTINE INSPECTION, THE DEVICE WAS FOUND TO DISPLAY THE WRENCH AND BATTERY ICONS AND FAILED TO POWER ON. THERE WAS NO PATIENT USE ASSOCIATED WITH THE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 LIFEPAK 1000 DEFIBRILLATOR MKJ PHYSIO-CONTROL, INC. 1000 NA

Patients

Seq Age Sex Outcome Treatment
1 NA