FDA Adverse Event Malfunction Summary report: N

SILICONE ULTRAVIOLET-ABSORBING POSTERIOR CHAMBER THREE PIECE FOLDABLE INTRAOCULA

MDR report key: 1903010 · Received November 11, 2010

Report

Report Number
2023826-2010-01122
Event Type
Malfunction
Date Received
November 11, 2010
Date of Event
October 14, 2010
Report Date
October 14, 2010
Manufacturer
STAAR SURGICAL CO.
Product Code
HQL
PMA / PMN Number
P900048
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
PHYSICIAN ASSISTANT

Narratives

Additional Manufacturer Narrative · 1

(B)(4): RESULTS - VISUAL INSPECTION OF THE RETURNED PRODUCT SHOWED THE OPTIC WAS TORN AND A PIECE OF THE OPTIC WAS TORN OFF WITH A HAPTIC AND MISSING. THERE WAS BROWN SURGICAL RESIDUE ON THE LENS. CONCLUSION - AN INVESTIGATION OF THE ROOT CAUSES OF LENS TEARS, SIMILAR TO THAT STATED IN THIS CLAIM, WERE ADDRESSED IN A CAPA OPENED IN JUNE 2005, (WHICH WAS SUBSEQUENTLY CLOSED). THE CAPA ADDRESSED DELIVERY SYSTEM ISSUES INCLUDING ALL STAGES OF MFG OF THE INJECTORS AND CARTRIDGES. PROCESSES WERE REVIEWED AND REVISED AS OPPORTUNITIES FOR IMPROVEMENT WERE REVEALED. THE CAPA ALSO ADDRESSED HANDLING ERRORS. ALL INJECTOR/CARTRIDGE DIRECTIONS FOR USE (DFU) HAVE BEEN MODIFIED TO ADD FURTHER CLARIFICATIONS TO INSTRUCT THE USERS IN THE PROPER DELIVERY TECHNIQUES THAT ARE EFFECTIVE, AND MINIMIZE THE POTENTIAL FOR DAMAGING THE LENS. BASED ON THE COMPLAINT HISTORY AND THE EVAL OF THE RETURNED PRODUCT POSSIBLE ROOT CAUSES FOR LENS TEARS INCLUDE BOTH DELIVERY SYSTEM ISSUES AND HANDLING ERRORS BY THE CUSTOMER. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE SURGEON INSERTED AN AQ2015A THREE PIECE SILICONE LENS AND THE LENS TORE UPON INSERTION. THE LENS WAS REMOVED AND ANOTHER SAME MODEL WAS IMPLANTED WITHOUT ANY PT INJURY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SILICONE ULTRAVIOLET-ABSORBING POSTERIOR CHAMBER THREE PIECE FOLDABLE INTRAOCULA INTRAOCULAR LENS HQL STAAR SURGICAL CO. AQ2015A NA

Patients

Seq Age Sex Outcome Treatment
1 UNK STAARVISC II, LOT NUMBER UNK| CARTRIDGE MODEL AQ CARTRIDGE-FP| INJECTOR MODEL MSI-PM, LOT NUMBER UNK