FDA Adverse Event Malfunction Summary report: N

SILICONE ULTRAVIOLET-ABSORBING POSTERIOR CHAMBER THREE PIECE FOLDABLE INTRAOCULA

MDR report key: 1902985 · Received November 8, 2010

Report

Report Number
2023826-2010-01119
Event Type
Malfunction
Date Received
November 8, 2010
Date of Event
October 6, 2010
Report Date
October 12, 2010
Manufacturer
STAAR SURGICAL CO.
Product Code
HQL
PMA / PMN Number
P900048
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4): RESULTS - VISUAL INSPECTION OF THE RETURNED PRODUCT FOUND THE LENS OPTIC TORN AND ONE HAPTIC BENT. THERE WAS EVIDENCE OF CLEAR SURGICAL RESIDUE. CONCLUSIONS - BASED ON THE COMPLAINT HISTORY AND THE EVAL OF THE RETURNED PRODUCT, POSSIBLE ROOT CAUSES FOR LENS TEARS INCLUDE BOTH DELIVERY SYSTEM ISSUES AND HANDLING ERRORS BY THE CUSTOMER. INVESTIGATION OF THE ROOT CAUSES OF LENS TEARS, WERE ADDRESSED IN A CAPA OPENED IN JUNE 2005 (WHICH WAS SUBSEQUENTLY CLOSED). THE CAPA ADDRESSED DELIVERY SYSTEM ISSUES INCLUDING ALL STAGES OF MFG OF THE INJECTORS AND CARTRIDGES. PROCESSES WERE REVIEWED AND REVISED AS OPPORTUNITIES FOR IMPROVEMENT WERE REVEALED. THE CAPA ALSO ADDRESSED HANDLING ERRORS. ALL INJECTOR/CARTRIDGE DIRECTIONS FOR USE (DFU) HAVE BEEN MODIFIED TO ADD FURTHER CLARIFICATIONS TO INSTRUCT THE USERS IN THE PROPER DELIVERY TECHNIQUES THAT ARE EFFECTIVE, AND MINIMIZE THE POTENTIAL FOR DAMAGING THE LENS. (B)(4).

Description of Event or Problem · 1

THE REPORTER INDICATED THE SURGEON INSERTED AN AQ5010V SILICONE THREE PIECE LENS, BUT THE LENS WAS BENT AND THE SURGEON WAS NOT HAPPY WITH THE HAPTICS. THE LENS WAS PARTIALLY INSERTED AND WAS REMOVED WITH NO PATIENT INJURY. A DIFFERENT LENS WAS IMPLANTED. ADD'L INFO HAS BEEN REQUESTED, BUT NONE HAS BEEN FORTHCOMING. IF ADD'L INFO BECOMES AVAILABLE, A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SILICONE ULTRAVIOLET-ABSORBING POSTERIOR CHAMBER THREE PIECE FOLDABLE INTRAOCULA INTRAOCULAR LENS HQL STAAR SURGICAL CO. AQ5010V NA

Patients

Seq Age Sex Outcome Treatment
1 56 YR CARTRIDGE MODEL AND LOT NUMBER UNKNOWN| VISCOELASTIC MODEL STAARVISC II - LOT NUMBER UNK| INJECTOR MODEL AND LOT NUMBER UNKNOWN