FDA Adverse Event
Other
Summary report: N
EVERCROSS PTA BALLOON CATHETER -6MM X 200MM, L =
MDR report key: 1902911
·
Received November 12, 2010
Report
- Report Number
- MW5018217
- Event Type
- Other
- Date Received
- November 12, 2010
- Date of Event
- November 1, 2010
- Report Date
- November 12, 2010
- Manufacturer
- EV3 ENDOVASCULAR INC.
- Product Code
- DQY
- Report Source
- Voluntary report
- Reporter Location
- IN, US
- Reporter Occupation
- OTHER
Narratives
Description of Event or Problem · 1
PHYSICIAN INSERTED BALLOON CATHETER INSIDE EXISTING STENT ON PATIENT. BALLOON WAS INFLATED TO EXPAND THE STENT. PHYSICIAN DEFLATED BALLOON AND ATTEMPTED TO REMOVE THE BALLOON CATHETER. DURING THE REMOVAL PROCESS, THE TIP OF THE BALLOON GOT CAUGHT ON THE STENT STRUTS AND THE TIP BROKE OFF. AFTER NUMEROUS UNSUCCESSFUL ATTEMPTS TO SNARE THE BROKEN TIP, THE PHYSICIAN DECIDED TO DEPLOY A SECOND STENT TO COVER THE BROKEN TIP AND OPEN UP THE LESION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | EVERCROSS PTA BALLOON CATHETER -6MM X 200MM, L = | BALLOON CATHETER | DQY | EV3 ENDOVASCULAR INC. | 8424083 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 73 YR | Other |