FDA Adverse Event
Malfunction
Summary report: N
ADJ PIN COLLET 2.0-3.2MM
MDR report key: 1902892
·
Received November 12, 2010
Report
- Report Number
- 1811755-2010-01686
- Event Type
- Malfunction
- Date Received
- November 12, 2010
- Date of Event
- October 15, 2010
- Report Date
- October 15, 2010
- Manufacturer
- STRYKER INSTRUMENTS KALAMAZOO
- Product Code
- KIJ
- PMA / PMN Number
- K972367
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WA, US
- Reporter Occupation
- UNKNOWN
Narratives
Additional Manufacturer Narrative · 1
THE DEVICE WAS RETURNED TO THE MFR FOR INVESTIGATION. BASED ON THE QUALITY INVESTIGATION, THE IMPREGLON COATING ON THE DEVICE WAS WORN OFF. THIS CONDITION COULD CAUSE THE DEVICE TO STICK, AND NOT ALLOW THE COLLET TO RELEASE THE WIRE.
Description of Event or Problem · 1
IT WAS REPORTED THAT PINS ARE GETTING STUCK IN THE DEVICE. THERE ARE NO ALLEGATIONS OF ADVERSE CONSEQUENCES ASSOCIATED WITH THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ADJ PIN COLLET 2.0-3.2MM | INSTRUMENT, SURGICAL, ORTHOPEDIC, DC-POWERED MOTOR | KIJ | STRYKER INSTRUMENTS KALAMAZOO |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |