FDA Adverse Event
Malfunction
Summary report: N
MICRO DRILL
MDR report key: 1902878
·
Received November 12, 2010
Report
- Report Number
- 1811755-2010-01660
- Event Type
- Malfunction
- Date Received
- November 12, 2010
- Date of Event
- October 20, 2010
- Report Date
- October 21, 2010
- Manufacturer
- STRYKER INSTRUMENTS KALAMAZOO
- Product Code
- ERL
- PMA / PMN Number
- K040300
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MD, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE HANDPIECE WAS REC'D AT THE MFR FOR EVAL, AND THE REPORTED CONDITION OF THE DEVICE OVERHEATING WAS DUPLICATED. BASED ON THE INVESTIGATION DETAILS, THE LIKELY CAUSE WAS A CORRODED MOTOR AND ROTOR, WHICH WERE EACH REPLACED ALONG WITH OTHER COMPONENTS. SERVICE REPAIRED AND RETURNED THE DEVICE TO THE CUSTOMER.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE HANDPIECE OVERHEATED DURING A ROUTINE MAINTENANCE VISIT AND IN A NON-STERILE ENVIRONMENT. THIS DID NOT OCCUR DURING A SURGICAL PROCEDURE. THERE WAS NO PT INVOLVEMENT. THERE WERE NO ADVERSE CONSEQUENCES REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | MICRO DRILL | SURGICAL INSTRUMENT MOTORS AND ACCESSORIES/ATTAC | ERL | STRYKER INSTRUMENTS KALAMAZOO |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |