FDA Adverse Event
Malfunction
Summary report: N
CORE MICRO DRILL
MDR report key: 1902877
·
Received November 12, 2010
Report
- Report Number
- 1811755-2010-01662
- Event Type
- Malfunction
- Date Received
- November 12, 2010
- Date of Event
- October 21, 2010
- Report Date
- October 21, 2010
- Manufacturer
- STRYKER INSTRUMENTS KALAMAZOO
- Product Code
- HBE
- PMA / PMN Number
- K032303
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- VA, US
- Reporter Occupation
- UNKNOWN
Narratives
Additional Manufacturer Narrative · 1
THE HANDPIECE WAS REC'D AT THE MFR FOR EVAL, AND THE REPORTED CONDITION OF THE DEVICE OVERHEATING WAS DUPLICATED. BASED ON THE INVESTIGATION DETAILS, THE LIKELY CAUSE WAS CORROSION ON THE MOTOR, ROTOR, AND PRELOAD BEARING. THOSE PARTS WERE EACH REPLACED ALONG WITH OTHER COMPONENTS. SERVICE REPAIRED AND RETURNED THE DEVICE TO THE CUSTOMER.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE HANDPIECE BEGAN OVERHEATING DURING A PODIATRY PROCEDURE. THE CASE WAS SUCCESSFULLY COMPLETED WITH ANOTHER DEVICE. THERE WERE NO ADVERSE CONSEQUENCES REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CORE MICRO DRILL | SURGICAL INSTRUMENT MOTORS AND ACCESSORIES/ATTAC | HBE | STRYKER INSTRUMENTS KALAMAZOO |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |