FDA Adverse Event Malfunction Summary report: N

ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM

MDR report key: 1902845 · Received November 12, 2010

Report

Report Number
2531779-2010-02251
Event Type
Malfunction
Date Received
November 12, 2010
Date of Event
October 14, 2010
Report Date
October 14, 2010
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE PUMP WAS RETURNED TO ANIMAS AND THE EVAL REVEALED THAT THE KEYPAD WAS PEELING. ALL KEYPAD BUTTONS ARE INTERMITTENTLY RESPONDING AND REQUIRED EXCESSIVE FORCE TO ACTIVATE. THERE WAS EVIDENCE OF KEYPAD ADHESIVE FOUND UNDER ALL KEY CONTACTS.

Description of Event or Problem · 1

THE PT REPORTED THAT THE DOWN BUTTON HAD TO BE PRESSED MULTIPLE TIMES TO ELICIT A RESPONSE FROM THE KEYPAD.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ONE TOUCH PING NA

Patients

Seq Age Sex Outcome Treatment
1 43 YR