FDA Adverse Event Malfunction Summary report: N

COLLEAGUE CX TRIPLE CHANNEL VOLUMETRIC INFUSION PUMP

MDR report key: 1902712 · Received November 18, 2010

Report

Report Number
6000001-2010-04992
Event Type
Malfunction
Date Received
November 18, 2010
Date of Event
June 16, 2010
Report Date
June 28, 2010
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FRN
PMA / PMN Number
K041191
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS RETURNED TO BAXTER AND IS CURRENTLY IN THE PROCESS OF BEING EVALUATED. A FOLLOW-UP REPORT WILL BE FILED UPON COMPLETION OF THE EVALUATION OR IF ANY ADDITIONAL DETAILS BECOME AVAILABLE.

Additional Manufacturer Narrative · 1

(B)(4). THIS DEVICE IS AN UNREMEDIATED COLLEAGUE PUMP WITH A USER INTERFACE MODULE SOFTWARE VERSION OF 5.01.00. THE ROOT CAUSE INVESTIGATION IS IN PROGRESS THROUGH (B)(4). EVALUATION SUMMARY: THE REPORTED CONDITION OF A COLLEAGUE INFUSION PUMP WITH FAILURE CODE 550:320:844:0000 WAS CONFIRMED IN THE PUMP'S EVENT HISTORY. THIS CONDITION WAS CAUSED BY A DEFECTIVE MAIN SPEAKER. THE MAIN SPEAKER WAS REPLACED TO CORRECT THIS CONDITION.

Description of Event or Problem · 1

THE FACILITY REPORTED A COLLEAGUE INFUSION PUMP WITH FAILURE CODE 550:320:844:0000. THE EVENT WAS REPORTED TO HAVE OCCURRED AFTER PATIENT THERAPY. THERE WAS NO PATIENT INJURY, MEDICAL INTERVENTION NECESSARY, OR ADVERSE REACTION IN ASSOCIATION WITH THIS EVENT. DURING PRODUCT EVALUATION, A BAXTER SERVICE TECHNICIAN CONFIRMED FAILURE CODE 550:320:844:0000, BUT ALSO FOUND THE PUMP DID NOT AUDIBLY ALARM FOR THIS CONDITION. NO ADDITIONAL INFORMATION IS AVAILABLE. THE USER INTERFACE MODULE SOFTWARE VERSION OF THIS PUMP IS CURRENTLY UNKNOWN.