FDA Adverse Event Malfunction Summary report: N

CX50 ULTRASOUND SYSTEM

MDR report key: 19023182 · Received April 2, 2024

Report

Report Number
3019216-2024-00048
Event Type
Malfunction
Date Received
April 2, 2024
Date of Event
March 14, 2024
Report Date
June 21, 2024
Manufacturer
PHILIPS ULTRASOUND, INC
Product Code
IYN
UDI-DI
00884838025295
PMA / PMN Number
K123754
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

A THOROUGH INVESTIGATION WITH A DEVICE ANALYSIS WAS PERFORMED AND THE CABLE DAMAGE WAS CONFIRMED. THE EXPOSED LIVE WIRE WOULD HAVE PRODUCED AN ELECTRIC SHOCK TO THE USER OR ANY METAL SURFACES OF THE CART IF IT CAME INTO CONTACT. THERE WAS A PUNCTURE TO THE CABLE IN WHICH THE INSULATION ON THE OUTER SHEATH WAS STRIPPED OFF AND THE LIVE WIRE WAS EXPOSED. IT WAS NOT POSSIBLE TO PROVIDE AN ACCURATE CAUSE OF THE DAMAGE.

Additional Manufacturer Narrative · 0

H10: ADDITIONAL INFORMATION WAS RECEIVED 30APR2024 REPORTING THE USER, WHO WAS SHOCKED, HAD NO SYMPTOMS OR CONSEQUENCES AND DID NOT SEEK MEDICAL ATTENTION. THIS EVENT DOES NOT MEET THE CRITERIA FOR A SERIOUS INJURY. H11: CORRECTIONS TO INITIAL REPORT: ADVERSE EVENT/PRODUCT PROBLEM: PRODUCT PROBLEM TYPE OF REPORTED COMPLAINT: PRODUCT PROBLEM PATIENT OUTCOME CODE GRID: NO CLINICAL SIGNS, SYMPTOMS OR CONDITIONS HEALTH IMPACT GRID: MINOR INJURY/ILLNESS/IMPAIRMENT

Description of Event or Problem · 0

A USER FACILITY REPORT BFARM 08301/24 WAS RECEIVED THAT REPORTED A SERIOUS INJURY ON 14MAR2024 IN WHICH A DOCTOR'S ASSISTANT PICKED UP THE CX50 ULTRASOUND SYSTEM FOR A PATIENT EXAM AND TOUCHED THE UNINSULATED SECTION OF THE POWER CORD. THE USER RECEIVED AN ELECTRICAL SHOCK TO THE RIGHT HAND. IT WAS UNSPECIFIED IF MEDICAL INTERVENTION WAS REQUIRED AND THE EXTENT OF THE PHYSICAL INJURY FROM THE SHOCK IS CURRENTLY UNKNOWN. PHILIPS HAS STARTED AN INVESTIGATION OF THIS COMPLAINT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1666049 CX50 ULTRASOUND SYSTEM SYSTEM, IMAGING, PULSED DOPPLER, ULTRASONIC IYN PHILIPS ULTRASOUND, INC 00884838025295

Patients

Seq Age Sex Outcome Treatment
1 NA Male Other