FDA Adverse Event Injury Summary report: N

PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM

MDR report key: 1902113 · Received November 10, 2010

Report

Report Number
2953144-2010-02792
Event Type
Injury
Date Received
November 10, 2010
Date of Event
October 21, 2010
Report Date
October 21, 2010
Manufacturer
ABBOTT VASCULAR - VASCULAR SOLUTIONS
Product Code
MGB
PMA / PMN Number
P960043
Removal / Correction Number
NA
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE IS EXPECTED TO BE RETURNED FOR EVALUATION. IT HAS NOT YET BEEN RECEIVED.

Additional Manufacturer Narrative · 1

(B)(4). THE PLUNGER AND NEEDLES, THE COMPLETE SUTURE, POSTERIOR NEEDLE TIP, ANTERIOR AND POSTERIOR CUFFS AND LINK MATERIAL WERE NOT RETURNED WHICH LIMITED THE INVESTIGATION. EVALUATION OF THE RETURNED DEVICE COMPONENTS FOUND NO DETECTED DAMAGE TO THE DEVICE HANDLE, GUIDE TUBE, GUIDE, FOOT OR SHEATH. BLOW THROUGH MARKS WERE PRESENT ON THE POSTERIOR FOOT INDICATING THE POSTERIOR CUFF HAD BEEN EJECTED FROM THE POSTERIOR FOOT. INSPECTION OF THE FOOT ASSEMBLY DID NOT DETECT ANY DAMAGE TO INDICATE A POSSIBLE CUFF MISS OR NEEDLE STRIKE OR POSSIBLE MALFUNCTION OF THE DEVICE. DURING TESTING, A PROXY PLUNGER/NEEDLE ASSEMBLY WAS INSERTED TO TEST FOR NEEDLE TRAJECTORY AND THE RESULTS WERE ACCEPTABLE. A POSSIBLE CAUSE FOR THE REPORTED CUFF MISS MAY HAVE BEEN NEEDLE DEFLECTION DURING PLUNGER DEPLOYMENT DUE TO INTERACTION WITH HUMAN TISSUE, AGGRESSIVE AND FAST DEPLOYMENT OF THE PLUNGER OR A FAILURE TO MAINTAIN A STABLE POSITION OF THE DEVICE WITH RESPECT TO THE TISSUE TRACT DURING NEEDLE DEPLOYMENT. BASED ON THE INVESTIGATION FINDINGS, THE ROOT CAUSE FOR THE REPORTED CUFF MISS COULD NOT BE DETERMINED. NO MANUFACTURING OR QUALITY ISSUES WERE DETECTED. REVIEW OF THE DEVICE HISTORY RECORD FOR THIS LOT DID NOT PRODUCE ANY FINDINGS RELEVANT TO THIS REPORT.

Description of Event or Problem · 1

IT WAS REPORTED THAT A PHYSICIAN TRAINED IN THE USE OF THE PROGLIDE DEVICE ATTEMPTED ARTERIOTOMY CLOSURE OF A FEMORAL ARTERY AFTER AN INTERVENTIONAL PROCEDURE. REPORTEDLY, WHEN THE NEEDLE PLUNGER WAS REMOVED, A CUFF MISS OCCURRED. THE PROGLIDE DEVICE WAS REMOVED AND HEMOSTASIS WAS ACHIEVED USING A SECOND PROGLIDE DEVICE. THERE WERE NO REPORTED ADVERSE PATIENT EFFECTS. NO ADDITIONAL INFORMATION WAS PROVIDED.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING THE PROCEDURE WHEN THE BALLOON OF THE STENT DELIVERY SYSTEM (SDS) WAS INFLATED TO DEPLOY THE STENT, IT WAS NOTICED THAT THE STENT HAD DISLODGED FROM THE BALLOON OUTSIDE THE ANATOMY. THERE WAS NO REPORTED PATIENT EFFECT. NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM SUTURE MEDIATED CLOSURE MGB ABBOTT VASCULAR - VASCULAR SOLUTIONS NA 920356H

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention HEPARIN