AUTOMATED PD SET W/CASSETTE4 PRONG
Report
- Report Number
- 1423500-2010-05908
- Event Type
- Malfunction
- Date Received
- November 18, 2010
- Date of Event
- October 31, 2010
- Report Date
- October 31, 2010
- Manufacturer
- BAXTER HEALTHCARE - MOUNTAIN HOME
- Product Code
- FKX
- PMA / PMN Number
- K923065
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- OTHER
Narratives
(B)(6). THE SAMPLE WAS DISCARDED. A FOLLOW-UP REPORT WILL BE SUBMITTED UPON THE COMPLETION OF BAXTER'S INVESTIGATION.
(B)(4). THIS COMPLAINT FOR A CONNECTION ISSUE WAS NOT CONFIRMED DUE TO LACK OF SAMPLE. THE LOT NUMBER IS UNKNOWN THEREFORE A BATCH REVIEW WAS NOT PERFORMED. THE ROOT CAUSE COULD NOT BE DETERMINED. RENAL QUALITY ENGINEERING, ALONG WITH PLANT MANUFACTURING AND QUALITY PERSONNEL, WILL CONTINUE TO WATCH FOR SIMILAR OCCURRENCE TRENDS AND WILL TAKE CORRECTIVE/PREVENTIVE ACTION AS APPROPRIATE.
A PATIENT CONTACTED GLOBAL TECHNICAL SERVICES (GTS) REGARDING ASSISTANCE WITH A SYSTEM ERROR 2240 WHILE USING THE HOMECHOICE (HC) DURING DRAIN 3. GTS EXPLAINED THAT A LARGE AMOUNT OF AIR HAD ENTERED INTO THE DISPOSABLE SET. GTS HAD THE PATIENT CYCLE POWER AND EXPLAINED THAT THERAPY HAD ENDED. GTS THEN EXPLAINED THE PATIENT WOULD NEED TO START OVER WITH NEW SUPPLIES OR FINISH THERAPY MANUALLY. PRODUCT SURVEILLANCE CONTACTED THE HOME PATIENT REGARDING THIS REPORT. THE PATIENT STATED THAT HE DID NOT KEEP THE SAMPLE BECAUSE HE FIGURED OUT THAT THE ADAPTER TO THE TRANSFER SET WAS NOT SCREWED IN TIGHT ENOUGH. THE PATIENT STATED THAT THERE WAS SOME FLUID WHERE THE CONNECTION WAS, BUT NOT ANYWHERE ELSE, SO HE FIGURED OUT THAT THE CONNECTION WAS NOT TIGHT. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | AUTOMATED PD SET W/CASSETTE4 PRONG | SYSTEM, PERITONEAL, AUTOMATIC DELIVERY | FKX | BAXTER HEALTHCARE - MOUNTAIN HOME |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 51 YR |