ONE STEP BUTTON¿
Report
- Report Number
- 3005099803-2010-04768
- Event Type
- Malfunction
- Date Received
- November 18, 2010
- Date of Event
- November 1, 2010
- Report Date
- November 1, 2010
- Manufacturer
- BOSTON SCIENTIFIC - SPENCER
- Product Code
- KGC
- PMA / PMN Number
- K910584
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- PHYSICIAN
Narratives
EXACT AGE OF PATIENT IS UNKNOWN HOWEVER, OVER 18 YEARS OLD. THE DEVICE HAS BEEN RECEIVED FOR ANALYSIS. UPON RECEIPT AND COMPLETION OF THE FAILURE ANALYSIS OF THE COMPLAINT DEVICE, IF THERE IS ANY FURTHER RELEVANT INFORMATION FROM THAT REVIEW, A SUPPLEMENTAL MEDWATCH WILL BE FILED.
THE PULLWIRE WAS THREADED THROUGH PERCUTANEOUS STOMA MEASURING DEVICE (PSMD) AND ATTACHED ONTO ONE STEP DELIVERY SYSTEM WIRE LOOP. A VISUAL EXAMINATION OF THE DEVICE REVEALED THE BLUE NYLON COATING OF THE PULLWIRE HAD BEEN PEELED AT THE INTERFACE BETWEEN THE PULLWIRE AND DELIVERY SYSTEM WIRE LOOP. THE NYLON COATING OF THE PULLWIRE WAS PEELED DOWN TO WIRE IN MULTIPLE AREAS. REMNANT OF THE PEELED NYLON COATING WAS MEASURED TO BE APPROXIMATELY 38 CENTIMETERS IN LENGTH. THE CONDITION OF THE RETURNED UNIT WAS CONSISTENT WITH THE COMPLAINT INCIDENT THAT THE DEVICE PULLWIRE COATING PEELED. THE EDGE OF THE PSMD HAS BEEN DETERMINED TO BE TOO SHARP THUS CATCHING ON THE NYLON COATING OF THE PULLWIRE AND CAUSING THE COATING TO PEEL AS THE PULLWIRE IS PULLED BACK THROUGH THE PSMD. THE DESIGN OF THE PSMD HAS BEEN DETERMINED TO BE THE MOST PROBABLE ROOT CAUSE OF THE FAILURE. FURTHER INVESTIGATION OF THIS ISSUE IS ONGOING. A REVIEW OF THE DEVICE HISTORY RECORD WAS PERFORMED AND REVEALED NO ISSUES RELATED TO THIS COMPLAINT. A LOT HISTORY SEARCH WAS PERFORMED AND FOUND NO OTHER COMPLAINTS AGAINST LOT NUMBER 13701285.
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A ONE STEP BUTTON INITIAL PLACEMENT GASTROSTOMY KIT WAS USED DURING A PERCUTANEOUS ENDOSCOPIC GASTROSTOMY PROCEDURE PERFORMED ON (B)(6), 2010. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, WHEN THE PHYSICIAN REMOVED THE PULL WIRE, FROM THE PATIENT'S STOMA SITE, IT WAS NOTED THE COATING FROM THE PULLWIRE WAS PEELED OFF. NONE OF THE COATING FELL INTO THE PATIENT. THE PROCEDURE WAS COMPLETED WITH THIS DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE GOOD.
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A ONE STEP BUTTON INITIAL PLACEMENT GASTROSTOMY KIT WAS USED DURING A PERCUTANEOUS ENDOSCOPIC GASTROSTOMY PROCEDURE PERFORMED ON (B)(6, 2010. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, WHEN THE PHYSICIAN REMOVED THE PULL WIRE, FROM THE PATIENT'S STOMA SITE, IT WAS NOTED THE COATING FROM THE PULLWIRE WAS PEELED OFF. NONE OF THE COATING FELL INTO THE PATIENT. THE PROCEDURE WAS COMPLETED WITH THIS DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE GOOD.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ONE STEP BUTTON¿ | TUBE, GASTRO-ENTEROSTOMY | KGC | BOSTON SCIENTIFIC - SPENCER | M00568500 | 13701285 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |