FDA Adverse Event
Malfunction
Summary report: N
DETECTO SCALE
MDR report key: 1901739
·
Received November 9, 2010
Report
- Report Number
- 1901739
- Event Type
- Malfunction
- Date Received
- November 9, 2010
- Date of Event
- November 7, 2010
- Report Date
- November 9, 2010
- Manufacturer
- CARDINAL HEALTH
- Product Code
- FRW
- Product Problem
- Yes
- Report Source
- User Facility report
- Reporter Location
- MO, US
- Reporter Occupation
- RISK MANAGER
Narratives
Description of Event or Problem · 1
BABY SCALE READINGS INACCURATE BY 5-10% WT DIFFERENCES. BIOMED HAS CALIBRATED AND ATTEMPTED TO REMEDY THE ISSUE, AND WHEN THE SCALE READS THEIR TEST USING A 10 POUND WEIGHT IT READS ACCURATELY, BUT WHEN WEIGHING A MOVING INFANT THE WEIGHTS ARE CONSISTENTLY OFF BY 5-10% OF THE ACTUAL WEIGHT.====================== HEALTH PROFESSIONAL'S IMPRESSION======================SCALE DOES NOT READ ACCURATELY AND CONSISTENTLY ESPECIALLY WITH INFANT MOVING INTERMITTENTLY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | DETECTO SCALE | SCALE, INFANT | FRW | CARDINAL HEALTH | 6745 | * |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | * | NO OTHER THERAPIES |