FDA Adverse Event Malfunction Summary report: N

DETECTO SCALE

MDR report key: 1901739 · Received November 9, 2010

Report

Report Number
1901739
Event Type
Malfunction
Date Received
November 9, 2010
Date of Event
November 7, 2010
Report Date
November 9, 2010
Manufacturer
CARDINAL HEALTH
Product Code
FRW
Product Problem
Yes
Report Source
User Facility report
Reporter Location
MO, US
Reporter Occupation
RISK MANAGER

Narratives

Description of Event or Problem · 1

BABY SCALE READINGS INACCURATE BY 5-10% WT DIFFERENCES. BIOMED HAS CALIBRATED AND ATTEMPTED TO REMEDY THE ISSUE, AND WHEN THE SCALE READS THEIR TEST USING A 10 POUND WEIGHT IT READS ACCURATELY, BUT WHEN WEIGHING A MOVING INFANT THE WEIGHTS ARE CONSISTENTLY OFF BY 5-10% OF THE ACTUAL WEIGHT.====================== HEALTH PROFESSIONAL'S IMPRESSION======================SCALE DOES NOT READ ACCURATELY AND CONSISTENTLY ESPECIALLY WITH INFANT MOVING INTERMITTENTLY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 DETECTO SCALE SCALE, INFANT FRW CARDINAL HEALTH 6745 *

Patients

Seq Age Sex Outcome Treatment
1 * NO OTHER THERAPIES