640G INSULIN PUMP MMT-1712K
Report
- Report Number
- 2032227-2024-149402
- Event Type
- Malfunction
- Date Received
- April 2, 2024
- Date of Event
- March 17, 2024
- Report Date
- September 3, 2024
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- UDI-DI
- 000000763000317157
- PMA / PMN Number
- EXEMPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- BR
- Reporter Occupation
- 003
Narratives
PUMP WAS RECEIVED WITH A CRACKED CASE-CORNER OF BELT CLIP RAILS NEAR THE BATTERY TUBE COMPARTMENT. PUMP WAS ALSO RECEIVED WITH A CRITICAL PUMP ERROR (OPEN BOOK IMAGE) ALARM. UNABLE TO PERFORM THE FUNCTIONAL TESTS, INCLUDING THE SELF TEST, SLEEP CURRENT MEASUREMENT, ACTIVE CURRENT MEASUREMENT, REWIND TEST, PRIME/SEATING TEST, BASIC OCCLUSION TEST, OCCLUSION TEST, FORCE SENSOR TEST AND DISPLACEMENT TEST DUE TO CRITICAL PUMP ERROR (OPEN BOOK IMAGE) ALARM. SUCCESSFULLY UTILIZED CREST AND THUS TO DOWNLOAD HISTORY FILES, TRACES AND COMLINK3 FILES. CRITICAL PUMP ERROR (OPEN BOOK IMAGE) ALARM AND PUMP ERROR 35 ALARM CONFIRMED IN THE FORMATTED HISTORY FILE ON (B)(6) 2024 11:27:14.000 AND ON (B)(6) 2024 11:37:01.000. PUMP WAS CUT OPEN TO PERFORM VISUAL INSPECTION AND FOUND CORROSION ON THE PCBA 1, PCBA 2 AND FORCE SENSOR. NO CORROSION OR MOISTURE DAMAGE FOUND ON THE MOTOR AND VIBRATOR ASSEMBLY NOTED. PLEASE SEE BELOW FOR PUMP ERROR(S)/ALARM(S) NOTED 2 DAYS PRIOR TO THE EVENT DATE 17-MAR-2024 IN THE FORMATTED HISTORY FILE. INSERT BATTERY ALARM WAS FOUND ON: (B)(6) 2024 11:35:38.000 TO ON (B)(6) 2024 12:04:35.000. LOW BATTERY ALERT WAS FOUND ON: (B)(6) 2024 11:30:01.000. FAILED BATTERY ALERT/BATTERY FAILED ALARM WAS FOUND ON: (B)(6) 2024 11:42:27.000 TO ON (B)(6) 2024 11:44:31.000. POWER MANAGEMENT GRAPH WAS SUCCESSFULLY GENERATED. THE POWER MANAGEMENT TOOL CONFIRMED THE UNLOADED VOLTAGE (UL VLITH) AND LOADED VOLTAGE (LOADED VLITH) WERE WITHIN SPEC RANGE. INSERT BATTERY ALARM WAS EXPECTED SINCE THE BATTERY WAS REMOVED FROM THE PUMP. UNABLE TO TEST FOR LOW BATTERY ALERT AND FAILED BATTERY ALERT/BATTERY FAILED ALARM DUE TO CRITICAL PUMP ERROR (OPEN BOOK IMAGE) ALARM. LOW BATTERY ALERT AND FAILED BATTERY ALERT/BATTERY FAILED ALARM WERE UNKNOWN. TEST P-CAP AND RESERVOIR LOCKED PROPERLY INTO RESERVOIR COMPARTMENT DURING TESTING. THE FOLLOWING WERE NOTED DURING VISUAL INSPECTION: A SCRATCH AND A SERIAL NUMBER LABEL FADING. COSMETIC DAMAGE WAS CONFIRMED AT THE START ON THE BELT CLIP RAILS OF THE PUMP DURING ANALYSIS. CRITICAL PUMP ERROR (OPEN BOOK IMAGE) ALARM CONFIRMED DUE TO FATAL ALARM PUMP ERROR 35. CRITICAL PUMP ERROR (OPEN BOOK IMAGE) ALARM AND PUMP ERROR 35 ALARM CONFIRMED DUE TO CORROSION ON THE PCBA 1, PCBA 2 AND FORCE SENSOR. PUMP EXPOSED TO MOISTURE WAS CONFIRMED. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.
THIS MDR RELATED TO THE PUERTO RICO MANUFACTURING SITE HAS BEEN ASSIGNED A MEDWATCH NUMBER FROM THE MEDTRONIC MINIMED NORTHRIDGE SITE, PER VARIANCE 5. SELECT PATIENT INFORMATION CANNOT BE PROVIDED DUE TO REGIONAL PRIVACY REGULATIONS. THE REPORTED DEVICE IS NOT MARKETED IN THE UNITED STATES, BUT IT IS A SAME/SIMILAR DEVICE TO ONE THAT IS MARKETED OUTSIDE THE UNITED STATES. CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.
IT WAS REPORTED TO MEDTRONIC MINIMED THAT THE CUSTOMER EXPERIENCED PUMP ERROR 35, DAMAGE (PHYSICAL OR COSMETIC), EXPOSED TO MOISTURE. THE CUSTOMER REPORTED NO ADVERSE EVENT. THE EVENT INVOLVED PRODUCT(S) MMT-1712K. TROUBLESHOOTING WAS PERFORMED. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. MMT-1712K WAS REQUESTED AND CUSTOMER RESPONSE WAS THE DEVICE WILL BE RETURNED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1361511 | 640G INSULIN PUMP MMT-1712K | PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR | OYC | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-1712K | HG5UWMQ | 000000763000317157 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |