FDA Adverse Event Malfunction Summary report: N

COLLEAGUE CX VOLUMETRIC INFUSION PUMP SINGLE CHANNEL

MDR report key: 1901706 · Received November 17, 2010

Report

Report Number
6000001-2010-04932
Event Type
Malfunction
Date Received
November 17, 2010
Date of Event
August 15, 2010
Report Date
October 25, 2010
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FRN
PMA / PMN Number
K063696
Removal / Correction Number
6000001-3/15/05-007-C
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

(B)(4).A FOLLOW-UP MEDWATCH REPORT WILL BE SUBMITTED WHEN THE EVALUATION RESULTS OR IF ADDITIONAL INFORMATION BECOMES AVAILABLE.

Additional Manufacturer Narrative · 1

(B)(4). A DEVICE HISTORY REVIEW HAS BEEN PERFORMED AND THERE WERE NO EXCEPTIONS NOTED DURING THE MANUFACTURING OF THIS PRODUCT. A TREND REVIEW HAS BEEN PERFORMED AND THERE HAVE BEEN SIMILAR REPORTS MADE TO BAXTER. THE ROOT CAUSE IS IN PROGRESS THROUGH (B)(4).

Additional Manufacturer Narrative · 1

(B)(4). DEVICE EVALUATION: THE REPORTED CONDITION WAS CONFIRMED. THE ASSIGNABLE CAUSE COULD NOT BE DETERMINED AT THIS TIME. THIS IS A STAY-IN UNIT AND WILL NOT BE REPAIRED AT THIS TIME. ADDITIONAL: A SERVICE HISTORY REVIEW HAS BEEN PERFORMED AND REVEALED THAT THIS DEVICE WAS NOT PREVIOUSLY SERVICED BY BAXTER PRIOR TO THIS EVENT. BAXTER HAS PERFORMED A DEVICE HISTORY REVIEW AND IT REVEALED THAT THERE WERE NO EXCEPTIONS ASSOCIATED WITH THE MANUFACTURING OF THIS PRODUCT.

Description of Event or Problem · 1

DURING BAXTER'S REVIEW OF THE EVENT HISTORY FOR ANOTHER REPORT OF A COLLEAGUE INFUSION PUMP, IT WAS DISCOVERED THAT FAILURE CODE 810:11 OCCURRED DURING INFUSION WHICH CAUSED AN INTERRUPTION DURING DELIVERY. THERE WAS NO ADVERSE EVENT, PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS REPORT. THERE IS NO FURTHER COMPLAINT INFORMATION AVAILABLE. THE USER INTERFACE MODULE MASTER SOFTWARE VERSION FOR THIS DEVICE IS 5.09.90, CATEGORIZED AS REMEDIATED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COLLEAGUE CX VOLUMETRIC INFUSION PUMP SINGLE CHANNEL PUMP, INFUSION FRN BAXTER HEALTHCARE - SINGAPORE

Patients

Seq Age Sex Outcome Treatment
1