FDA Adverse Event Injury Summary report: N

PERCLOSE PROSTYLE SUTURE-MEDIATED CLOSURE AND REPAIR SYSTEM

MDR report key: 19016319 · Received April 1, 2024

Report

Report Number
2024168-2024-04035
Event Type
Injury
Date Received
April 1, 2024
Date of Event
March 8, 2024
Report Date
May 20, 2024
Manufacturer
ABBOTT VASCULAR
Product Code
MGB
UDI-DI
08717648344039
PMA / PMN Number
P960043
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KS
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

MANUFACTURER'S INVESTIGATION IS STILL PENDING AT THIS TIME. RESULTS AND CONCLUSIONS WILL BE PROVIDED IN THE FINAL REPORT. THE ADDITIONAL PERCLOSE PROSTYLE DEVICE REFERENCED IS FILED UNDER A SEPARATE MEDWATCH REPORT NUMBER.

Additional Manufacturer Narrative · 0

THE DEVICE WAS RETURNED FOR ANALYSIS. THE REPORTED SUTURE RETRIEVAL ISSUE WAS CONFIRMED AS A LINK SEPARATION LOT HISTORY RECORD (LHR) AND CORRECTIVE AND PREVENTIVE ACTIONS (CAPA) REVIEWS WERE PERFORMED AND REVEALED NO INDICATION OF A PRODUCT QUALITY ISSUE. THE INVESTIGATION DETERMINED THAT THE OBSERVED LINK SEPARATION AND UNEXPECTED MEDICAL INTERVENTION APPEAR TO BE RELATED TO OPERATIONAL CONTEXT. IT IS LIKELY THAT AN INTERACTION OF THE DEVICE WITH PATIENT ANATOMY OR LINK PULLED UNTIL IT BREAKS DUE TO CIRCUMSTANCES OF THE PROCEDURE. BASED ON THE INFORMATION REVIEWED, THERE IS NO INDICATION OF A PRODUCT QUALITY ISSUE WITH RESPECT TO THE DESIGN, MANUFACTURE, OR LABELING OF THE DEVICE. CORRECTION: D4 - LOT #: UPDATED FROM 3032841 TO 3040442. D4 - EXPIRATION DATE: UPDATED. D4 - PRIMARY UDI NUMBER: UPDATED. H4 - DEVICE MFG DATE: UPDATED.

Description of Event or Problem · 0

IT WAS REPORTED THAT THIS WAS A CLOSURE OF A RIGHT COMMON FEMORAL ARTERY USING TWO PROSTYLE DEVICES AFTER AN INTERVENTIONAL PERCUTANEOUS CORONARY INTERVENTION PROCEDURE WITH A 6FR SHEATH. REPORTEDLY, A CUFF MISS [SUTURE RETRIEVAL ISSUE] OCCURRED WITH BOTH DEVICES. MANUAL COMPRESSION WAS USED TO ACHIEVE HEMOSTASIS. THERE WAS NO ADVERSE PATIENT SEQUELA AND NO REPORTED CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE OR THERAPY. NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1767512 PERCLOSE PROSTYLE SUTURE-MEDIATED CLOSURE AND REPAIR SYSTEM SUTURE MEDIATED CLOSURE MGB ABBOTT VASCULAR 3040442 08717648344039

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown Required Intervention