FDA Adverse Event Malfunction Summary report: N

ANIMAS INSULIN INFUSION PUMP

MDR report key: 1901626 · Received November 10, 2010

Report

Report Number
2531779-2010-02226
Event Type
Malfunction
Date Received
November 10, 2010
Report Date
October 13, 2010
Manufacturer
ANIMAS CORP.
Product Code
LZG
PMA / PMN Number
K042873
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SW
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE PUMP WAS RETURNED TO ANIMAS FOR EVALUATION. THE RESULTS OF THE INVESTIGATION WERE AS NOTED: THE KEYPAD APPEARED TO BE INTACT (NO LIFTING OR PEELING), THE PUMP WAS INTERMITTENTLY RESPONDING TO KEYPAD BUTTON PRESSES, THE UP KEY CONTACT WAS MISALIGNED, THE UP AND DOWN KEY CONTACT WERE INVERTED AND MISALIGNED AND DID NOT ALWAYS SPRING BACK, AND THERE WAS EVIDENCE OF ADHESIVE/CONTAMINATION UNDER ALL THE KEY CONTACTS.

Description of Event or Problem · 1

ACCORDING TO THE DISTRIBUTOR, THE PT HAS TO PRESS THE BUTTONS SEVERAL TIMES FOR THE PUMP TO RESPOND.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ANIMAS INSULIN INFUSION PUMP INSULIN INFUSION PUMP LZG ANIMAS CORP. ANIMAS 2020 NA

Patients

Seq Age Sex Outcome Treatment
1