FDA Adverse Event Malfunction Summary report: N

ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM

MDR report key: 1901623 · Received November 10, 2010

Report

Report Number
2531779-2010-02223
Event Type
Malfunction
Date Received
November 10, 2010
Report Date
October 12, 2010
Manufacturer
ANIMAS CORP.
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CO, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE PUMP WAS RETURNED TO ANIMAS FOR EVALUATION. THE RESULTS OF THE INVESTIGATION WERE AS NOTED: THE KEYPAD APPEARED TO BE SWOLLEN, THE PUMP WAS INTERMITTENTLY RESPONDING TO KEYPAD BUTTON PRESSES, THE BUTTONS REQUIRED EXCESSIVE FORCE TO ACTIVATE, THE UP KEY CONTACT WAS MISALIGNED, THE CONTRAST KEY CONTACT WAS INVERTED AND MISALIGNED, AND THERE WAS EVIDENCE OF ADHESIVE/CONTAMINATION UNDER THAT ALL KEY CONTACTS.

Description of Event or Problem · 1

IT WAS REPORTED THAT PUMP DOES NOT ALWAYS RESPOND TO THE KEYPAD BUTTON PRESSES. THE PUMP REPORTED HAS NOT BEEN EXPOSED TO WATER AND THE KEYPAD IS INTACT. THERE WAS NO ADVERSE EVENT ASSOCIATED WITH THIS COMPLAINT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORP. ONE TOUCH PING NA

Patients

Seq Age Sex Outcome Treatment
1