FDA Adverse Event Injury Summary report: N

TALENT STENT GRAFT SYSTEM WITH XCELERANT - HYDRO

MDR report key: 1901597 · Received November 12, 2010

Report

Report Number
2953200-2010-02218
Event Type
Injury
Date Received
November 12, 2010
Date of Event
October 14, 2010
Report Date
October 14, 2010
Manufacturer
MEDTRONIC CARDIOVASCULAR
Product Code
MIH
PMA / PMN Number
P070027
Removal / Correction Number
NA
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
NJ, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4). EVAL RESULTS AND CONCLUSION: (SEVERELY TORTUOUS ILIAC ARTERIES).

Description of Event or Problem · 1

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

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TALENT STENT GRAFT SYSTEM WITH XCELERANT - HYDRO MIH MEDTRONIC CARDIOVASCULAR NA V00238332

Patients

Seq Age Sex Outcome Treatment
1 75 YR Required Intervention