FDA Adverse Event Injury Summary report: N

ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM

MDR report key: 1901328 · Received November 13, 2010

Report

Report Number
2531779-2010-02306
Event Type
Injury
Date Received
November 13, 2010
Date of Event
October 14, 2010
Report Date
November 13, 2010
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. THE ANIMAS CDE INVOLVED WITH THIS CONTACT INDICATED THAT THE PATIENT WAS CONTINUING TO USE THE PUMP. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE DRAWN AT THIS TIME.

Additional Manufacturer Narrative · 1

(B)(4). DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2012 WITH THE FOLLOWING FINDINGS: THERE WAS NO DATA IN THE BLACK BOX OR DOWNLOAD HISTORY FROM THE TIME OF THE REPORTED HOSPITALIZATION DUE TO CONTINUED PATIENT USE. A REVIEW OF THE TOTAL DAILY DOSE HISTORY SHOWED THAT THE DAILY INSULIN DELIVERY TOTALS CORRECTLY REFLECTED THE USER PROGRAMMED BASAL RATES. THERE WAS NO ACTIVITY OUTSIDE NORMAL USE OBSERVED IN THE BLACK BOX OR DOWNLOAD HISTORY. A 29 HOUR FLOW ACCURACY TEST WAS PERFORMED AND THE PUMP WAS FOUND TO BE DELIVERING WITHIN SPECIFICATIONS.

Description of Event or Problem · 1

THE PATIENT AND HIS WIFE CONTACTED ANIMAS ALLEGING THAT SHE WAS ADMITTED TO A HOSPITAL FOR ELEVATED BLOOD GLUCOSE LEVELS AND TREATED WITH IV INSULIN. THE PATIENT CLAIMED THAT HE PERFORMED AN INSERTION SITE CHANGE ON (B)(6), 2010, AND WENT TO BED WITH NORMAL BLOOD GLUCOSE. THE PATIENT CLAIMED THAT HE WOKE UP THE NEXT MORNING VOMITING AND WITH A BLOOD GLUCOSE LEVEL OF 552 MG/DL AT 1:00 PM. HE RETESTED AT 2:00 PM AND SEVERAL TIMES UNTIL 5:00 PM AND GOT "HI" EACH TIME. THE PATIENT WAS ADMITTED TO A HOSPITAL WITH A BLOOD GLUCOSE LEVEL OF 656 MG/DL.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ONE TOUCH PING NA

Patients

Seq Age Sex Outcome Treatment
1 53 YR Hospitalization| L| R