FDA Adverse Event Malfunction Summary report: N

TAXUS LIBERTE PACLITAXEL-ELUTING CORONARY STENT SYSTEM

MDR report key: 1901309 · Received November 17, 2010

Report

Report Number
2134265-2010-05365
Event Type
Malfunction
Date Received
November 17, 2010
Date of Event
October 20, 2010
Report Date
October 22, 2010
Manufacturer
BOSTON SCIENTIFIC - GALWAY
Product Code
NIQ
PMA / PMN Number
P060008
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS NOT RETURNED FOR ANALYSIS. THE MANUFACTURING BATCH RECORD REVIEW CONFIRMED THAT THE DEVICE MET ALL MATERIAL, ASSEMBLY AND PERFORMANCE SPECIFICATIONS. THE MOST PROBABLE ROOT CAUSE IS OPERATIONAL CONTEXT AS DEVICE PERFORMANCE WAS LIMITED DUE TO ANATOMICAL/PROCEDURAL FACTORS. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A STENTING TREATMENT PROCEDURE THE STENT WAS PARTIALLY DEPLOYED. THE 99% STENOSED TARGET LESION WAS LOCATED IN THE OSTIAL CORONARY ARTERY. A 3.50X16MM TAXUS LIBERTE STENT DELIVERY SYSTEM (SDS) WAS ADVANCED TO THE TARGET LESION AND WHEN THE PHYSICIAN ATTEMPTED TO DEPLOY THE STENT AT 18ATMS THE STENT ONLY PARTIALLY DEPLOYED. IT WAS NOTED THAT THE DEVICE WAS SUCCESSFULLY REMOVED FROM THE PATIENT AND THE PROCEDURE WAS COMPLETED WITH A 3.5X13MM NON BSC STENT. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT'S CONDITION IS STABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TAXUS LIBERTE PACLITAXEL-ELUTING CORONARY STENT SYSTEM CORONARY DRUG-ELUTING STENT NIQ BOSTON SCIENTIFIC - GALWAY H7493894016350 12975301

Patients

Seq Age Sex Outcome Treatment
1