FDA Adverse Event Injury Summary report: N

DEXTRUS 4137

MDR report key: 1901279 · Received November 9, 2010

Report

Report Number
1028232-2010-02395
Event Type
Injury
Date Received
November 9, 2010
Date of Event
September 3, 2010
Report Date
October 18, 2010
Manufacturer
BIOTRONIK SE & CO. KG
Product Code
NVZ
PMA / PMN Number
P950037
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE WAS NOT RETURNED FOR ANALYSIS. THE ANALYSIS IS, THEREFORE, BASED ON THE INSPECTION OF THE QUALITY DOCUMENTS ACCOMPANYING THIS PARTICULAR DEVICE. THE MFG PROCESS FOR THIS DEVICE WAS RE-INVESTIGATED. ALL PRODUCTION STEPS HAD BEEN PERFORMED ACCORDINGLY. THERE WAS NO SIGN OF ANY INCONSISTENCY DURING THE MFG PROCESS, WHICH MIGHT BE RELATED TO THE CLINICAL OBSERVATION. IN SUMMARY, THE DEVICE WAS NOT RETURNED FOR ANALYSIS. THE REVIEW OF THE QUALITY DOCUMENTS CONFIRMED A REGULAR DEVICE MFG.

Description of Event or Problem · 1

BOSTON SCIENTIFIC CRM RECEIVED INFO THAT THE SENSING ON THE RIGHT VENTRICULAR LEAD DECREASED FROM 7MV TO 1.6MV. IT WAS NOTED THAT THE PHYSICIAN STUCK THE VEIN MORE MEDIAL THAN USUAL, RESULTING IN THE LEAD HAVING A BEND AT THE ACCESS POINT. THIS LEAD MICRO-DISLODGED. THEREFORE, THE LEAD WAS SUCCESSFULLY REPOSITIONED. NO ADVERSE PT EFFECTS WERE REPORTED. THIS LEAD REMAINS IMPLANTED. NO ADDITIONAL INFO IS AVAILABLE AT THIS TIME. IF ADDITIONAL INFO BECOMES AVAILABLE, THIS REPORT WILL BE UPDATED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 DEXTRUS 4137 PACER LEAD NVZ BIOTRONIK SE & CO. KG 358764

Patients

Seq Age Sex Outcome Treatment
1 UNK Hospitalization