FDA Adverse Event Malfunction Summary report: N

PD CYCLER 110 VOLT HOME CHOICEPRO AUTOMATED

MDR report key: 1901268 · Received November 17, 2010

Report

Report Number
1423500-2010-05860
Event Type
Malfunction
Date Received
November 17, 2010
Date of Event
September 20, 2010
Report Date
October 25, 2010
Manufacturer
BAXTER HEALTHCARE - LARGO
Product Code
FKX
PMA / PMN Number
K053512
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS RECEIVED AND EVALUATED. A REVIEW OF THE DEVICE LOGS REVEALED AN INCREASED INTRAPERITONEAL VOLUME (IIPV). THE ASSIGNABLE CAUSE OF THE IIPV WAS DETERMINED TO BE: INSUFFICIENT DRAIN - ONE OR MORE CYCLES ADVANCES TO NEXT FILL WHEN SLOW / NO FLOW OCCURRED ABOVE THE MINIMUM DRAIN VOLUME THRESHOLD. A SERVICE HISTORY REVIEW WAS PERFORMED. NO ISSUES WERE IDENTIFIED THAT MAY HAVE CONTRIBUTED TO THE IIPV. BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. THE ROOT CAUSE INVESTIGATION IS IN PROGRESS THROUGH (B)(4).

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE HAS BEEN RECEIVED, AND THE EVALUATION IS IN PROCESS. A FOLLOW-UP MDR WILL BE SUBMITTED UPON COMPLETION OF THE EVALUATION OR IF ANY ADDITIONAL INFORMATION IS RECEIVED.

Description of Event or Problem · 1

DURING INITIAL ASSESSMENT OF A RETURNED HOMECHOICE MACHINE, A BAXTER TECHNICIAN FOUND AN INCREASED INTRAPERITONEAL VOLUME (IIPV) SITUATION WHICH OCCURRED ON (B)(6) 2010 DURING DRAIN CYCLE 12. THE ULTRAFILTRATION VOLUME WAS 65 ML (MILLILITERS). THE PROGRAMMED FILL VOLUME WAS 200 ML. THIS EVENT MEETS OVERFILL CRITERIA. THIS IS REPORT 7 OF 7. ON (B)(6) 2010, PRODUCT SURVEILLANCE FOLLOWED UP WITH THE NURSE, AND PROVIDED THE RESULTS OF EVALUATION. THE NURSE STATED THAT THE DEVICE WAS USED WHILE THE PATIENT WAS IN THE HOSPITAL. THIS IS NOT THE PATIENT'S CLINIC AND NO ADDITIONAL INFORMATION WAS OBTAINED FROM THE NURSE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PD CYCLER 110 VOLT HOME CHOICEPRO AUTOMATED SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - LARGO

Patients

Seq Age Sex Outcome Treatment
1