FDA Adverse Event Malfunction Summary report: N

# 9 SCORPIO 4:1 CERAMIC BLOCKS

MDR report key: 1901132 · Received November 10, 2010

Report

Report Number
2249697-2010-01517
Event Type
Malfunction
Date Received
November 10, 2010
Date of Event
October 18, 2010
Report Date
October 18, 2010
Manufacturer
STRYKER ORTHOPAEDICS MAHWAH
Product Code
LXH
PMA / PMN Number
NA
Removal / Correction Number
2249697-12/14/09-012R
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
EI
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

METHOD: COMPLAINT HISTORY REVIEW. EVALUATION SUMMARY: A REVIEW OF THE PRODUCT EXPERIENCE REPORTING DATABASE AND COMPLAINT FILES SHOW THAT A TREND HAS BEEN PREVIOUSLY IDENTIFIED REGARDING THE CERAMIC RAILS FRACTURING AND/OR FALLING OUT OF THE CUTTING BLOCKS. THE RESULTS OF THE INVESTIGATION PERFORMED INDICATED THAT THIS EVENT IS RELATED TO A TREND OF SIMILAR EVENTS WHERE CERAMIC RAILS HAVE FRACTURED. THE RESULTS OF THE INVESTIGATION CONCLUDED THAT THE ROOT CAUSE OF THIS TREND IS DESIGN RELATED. CONTRIBUTING FACTORS TO THIS ROOT CAUSE INCLUDE THE DIAMETER AND PROXIMITY OF THE CERAMIC RAIL TO THE ANTERIOR AND POSTERIOR FACES OF THE CUTTING BLOCK, AS WELL AS THE ELASTIC PROPERTIES OF ALUMINIUM, COMPARED TO THAT OF CERAMIC.

Description of Event or Problem · 1

IT WAS REPORTED: "THE CUSTOMER, [...] REPORTED VIA OUR DISTRIBUTOR, (B)(6) IN (B)(6) 2010 THAT THE UNITS FAILED UNDER (B)(4) WHEN THE CUSTOMER INSPECTED." TWO DEVICES CATALOG # 8000-0009 (UNKNOWN LOT#) WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 # 9 SCORPIO 4:1 CERAMIC BLOCKS INSTRUMENT LXH STRYKER ORTHOPAEDICS MAHWAH NA UNK

Patients

Seq Age Sex Outcome Treatment
1 NA Other